Decision Summary

Upload-ready calibration protocol overview
Instrument P115s1
Session 34
Location Rome-ISAC
Calibration person Hanna Rohringer
CF file Pandora115s1_CF_v34d20211213.txt
Software version 1.8.89
Calibration vector calV14
Reference day 20230426
Overall status development
O3Ext: YesO3Synt: YesNO2: YesSO2: With CareHCHO: With CareH2O: Yes
Instrument P115s1
Session 34
Events 357
Figures 17
Action time 6h 16m

Action Timeline

111 action intervals
Mar 09, 2026Mar 17, 202609:07 AMCreateReferenceWavelengthCalibrationFieldCalibrationAssignCalFileCreateReference 36s finishedCreateReference 0s abortedCreateReference 21s finishedCreateReference 0s abortedCreateReference 18s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 14s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 25m 31s finishedCreateReference 0s abortedCreateReference 2m 37s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 3m 31s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 1m 46s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 1m 17s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 1m 24s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 4s abortedCreateReference 27s finishedCreateReference 0s abortedCreateReference 30s abortedWavelengthCalibration 3s abortedWavelengthCalibration 8s finishedCreateReference 2m 43s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 24s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 50s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 26s finishedCreateReference 0s abortedCreateReference 56s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 40s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 40s abortedCreateReference 1m 50s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 50s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 35s finishedCreateReference 0s finishedCreateReference 3s abortedFieldCalibration 2m 8s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 1h 20m finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s abortedFieldCalibration 3h 33m finishedFieldCalibration 0s abortedFieldCalibration 1m 6s abortedFieldCalibration 7m 34s finishedFieldCalibration 0s abortedFieldCalibration 4s finishedFieldCalibration 0s finishedFieldCalibration 0s abortedFieldCalibration 6m 26s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 4m 2s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedAssignCalFile 36s finishedAssignCalFile 29s finishedFieldCalibration 6m 56s abortedFieldCalibration 1m 55s abortedFieldCalibration 1m 54s aborted

Duration by Step

FieldCalibration5h 26mCreateReference48m 35sAssignCalFile1m 5sWavelengthCalibration11s

Action Summary

Action Events Runs Aborted Time
AssignCalFile 6 2 0 1m 5s
CreateReference 162 71 23 48m 35s
FieldCalibration 169 36 8 5h 26m
No action 15 0 0 0s
WavelengthCalibration 5 2 1 11s

Session Figures

17 figures from Google Drive/figures
calprotocol section

Wavelength Calibration

Slit function parameters

Pandora115s1_WavelengthCalibration_SlitFuncParam.jpeg - 165 KB
Slit function parameters
reference period

Reference Day Decision Support

Upload the FOV and L0 raw-count evidence used to justify the selected reference day before reviewing the CreateReference figures.

calprotocol section

Reference Period and CreateReference

Create reference L1CHARFILT

Pandora115s1_CreateReference_L1CHARFILT_OPEN_U340.jpeg - 157 KB
Create reference L1CHARFILT

Create reference SELSPECS

Pandora115s1_CreateReference_SELSPECS_OPEN_U340.jpeg - 129 KB
Create reference SELSPECS

Create reference SLERR (STRAY)

Pandora115s1_CreateReference_SLERR(STRAY)_OPEN_U340.jpeg - 177 KB
Create reference SLERR (STRAY)

Create reference AODEST

Pandora115s1_CreateReference_AODEST.jpeg - 64 KB
Create reference AODEST

Create reference SENSCORRFIN (SENS)

Pandora115s1_CreateReference_SENSCORRFIN(SENS)_OPEN_U340.jpeg - 121 KB
Create reference SENSCORRFIN (SENS)

Create reference DISPFIT (DISP)

Pandora115s1_CreateReference_DISPFIT(DISP).jpeg - 101 KB
Create reference DISPFIT (DISP)

Create reference WLCHANGE (DISP)

Pandora115s1_CreateReference_WLCHANGE(DISP)_OPEN_U340.jpeg - 212 KB
Create reference WLCHANGE (DISP)

Create reference SHIFT (CLOS)

Pandora115s1_CreateReference_SHIFT(CLOS)_OPEN_U340.jpeg - 139 KB
Create reference SHIFT (CLOS)
field calibration

O3

FieldCalibration VC O3

Pandora115s1_FieldCalibration_VC_O3.jpeg - 101 KB
FieldCalibration VC O3

FieldCalibration SCREFINT O3

Pandora115s1_FieldCalibration_SCREFINT_O3.jpeg - 75 KB
FieldCalibration SCREFINT O3
field calibration

NO2

NO2 E-MLE settings applied fw1out=(); fw2out=(); dqf2in=(); ddlim=(-1.000,-1.000); shilim(nm)=(-0.050,0.050); uscmax=4.348e-04; wrmsmax=2.467e+00; atvarmax(%)=2.000e+01; amflim=(1.200,5.000); amflimvar=(0.000,1.000); perc=50.00; ngroup=30; timeaver(min)=0.00; redu=2; scref=(2.506e-04,2.199e-04,2.335e-04); uscref=(1.205e-05,8.913e-06,1.043e-05)

FieldCalibration MLE NO2

Pandora115s1_FieldCalibration_MLE_NO2.jpeg - 89 KB
FieldCalibration MLE NO2

FieldCalibration VCMLE NO2

Pandora115s1_FieldCalibration_VCMLE_NO2.jpeg - 150 KB
FieldCalibration VCMLE NO2
field calibration

SO2

SO2 MLE settings applied fw1out=(); fw2out=(); dqf2in=(); ddlim=(-1.000,-1.000); shilim(nm)=(-0.050,0.050); uscmax=1.104e-02; wrmsmax=1.347e+00; atvarmax(%)=2.000e+01; amflim=(1.000,2.200); amflimvar=(0.000,1.000); perc=0.00; ngroup=8; timeaver(min)=0.00; redu=0; scref=(6.086e-05); uscref=(3.156e-05)

FieldCalibration MLE SO2

Pandora115s1_FieldCalibration_MLE_SO2.jpeg - 142 KB
FieldCalibration MLE SO2

FieldCalibration VCMLE SO2

Pandora115s1_FieldCalibration_VCMLE_SO2.jpeg - 124 KB
FieldCalibration VCMLE SO2
field calibration

HCHO

HCHO E-MLE settings applied fw1out=(); fw2out=(); dqf2in=(103); ddlim=(-1.000,-1.000); shilim(nm)=(-0.050,0.050); uscmax=1.303e-02; wrmsmax=5.867e-01; atvarmax(%)=4.000e+01; amflim=(1.000,7.000); amflimvar=(0.000,1.000); perc=50.00; ngroup=8; timeaver(min)=0.00; redu=2; scref=(1.631e-04); uscref=(1.931e-05)

FieldCalibration MLE HCHO

Pandora115s1_FieldCalibration_MLE_HCHO.jpeg - 57 KB
FieldCalibration MLE HCHO

FieldCalibration VCMLE HCHO

Pandora115s1_FieldCalibration_VCMLE_HCHO.jpeg - 182 KB
FieldCalibration VCMLE HCHO

Automatically harvested metadata

Automatically Harvested Metadata

Manual review and decision

Manual Review and Decision

Calibration notes, data-quality findings, and whole-period evaluation.

Data Evaluation

Could there be an offset between VPs?

No

Are there breakpoint-analysis plots or suspicious discontinuities?

No

Do daily figures before and after the new VP look consistent?

No

Is there an AMF dependency for one or more VPs?

No

Is there seasonality, for example negative values in winter?

No

Does the daily shape follow the trop data roughly?

No

If there is a discrepancy, what are the DS and Sky viewing directions?

No

Is there a co-located or nearby instrument for comparison?

No

O3Ext

Decision for O3Ext
Yes
looks pretty ok

O3Synt

Decision for O3Synt
Yes

NO2

Decision for NO2
Yes
looks good, E-MLE used

SO2

Decision for SO2
With Care
there is a data jump but its lower the 1e-4 thats fine

HCHO

Decision for HCHO
With Care
there is a data jump with in VP because of res change. It will need a 19 calibration a a new VP there to account for that

H2O

Decision for H2O
Yes
ok