Decision Summary

Upload-ready calibration protocol overview
Instrument P115s1
Session 29
Location -
Calibration person HRohringer
CF file -
Software version 1.8.89
Calibration vector -
Reference day -
Overall status development
O3Ext: With CareO3Synt: With CareNO2: YesSO2: UnusableHCHO: UnusableH2O: Yes
Instrument P115s1
Session 29
Events 0
Figures 15
Action time 0s

Action Timeline

No start/finish intervals found.
No timeline intervals found

Duration by Step

No completed durations found

Action Summary

Action Events Runs Aborted Time
No actions loaded

Session Figures

15 figures from Google Drive/figures
calprotocol section

Wavelength Calibration

Slit function parameters

Pandora115s1_WavelengthCalibration_SlitFuncParam.jpeg - 159 KB
Slit function parameters
reference period

Reference Day Decision Support

Upload the FOV and L0 raw-count evidence used to justify the selected reference day before reviewing the CreateReference figures.

calprotocol section

Reference Period and CreateReference

Create reference L1CHARFILT

Pandora115s1_CreateReference_L1CHARFILT_OPEN_U340.jpeg - 156 KB
Create reference L1CHARFILT

Create reference SELSPECS

Pandora115s1_CreateReference_SELSPECS_OPEN_U340.jpeg - 131 KB
Create reference SELSPECS

Create reference SLERR (STRAY)

Pandora115s1_CreateReference_SLERR(STRAY)_OPEN_U340.jpeg - 129 KB
Create reference SLERR (STRAY)

Create reference AODEST

Pandora115s1_CreateReference_AODEST.jpeg - 63 KB
Create reference AODEST

Create reference SENSCORRFIN (SENS)

Pandora115s1_CreateReference_SENSCORRFIN(SENS)_OPEN_U340.jpeg - 113 KB
Create reference SENSCORRFIN (SENS)

Create reference DISPFIT (DISP)

Pandora115s1_CreateReference_DISPFIT(DISP).jpeg - 113 KB
Create reference DISPFIT (DISP)

Create reference WLCHANGE (DISP)

Pandora115s1_CreateReference_WLCHANGE(DISP)_OPEN_U340.jpeg - 191 KB
Create reference WLCHANGE (DISP)

Create reference SHIFT (CLOS)

Pandora115s1_CreateReference_SHIFT(CLOS)_OPEN_U340.jpeg - 170 KB
Create reference SHIFT (CLOS)
field calibration

O3

FieldCalibration VC O3

Pandora115s1_FieldCalibration_VC_O3.jpeg - 102 KB
FieldCalibration VC O3

FieldCalibration SCREFINT O3

Pandora115s1_FieldCalibration_SCREFINT_O3.jpeg - 65 KB
FieldCalibration SCREFINT O3
field calibration

NO2

FieldCalibration MLE NO2

Pandora115s1_FieldCalibration_MLE_NO2.jpeg - 162 KB
FieldCalibration MLE NO2

FieldCalibration VCMLE NO2

Pandora115s1_FieldCalibration_VCMLE_NO2.jpeg - 107 KB
FieldCalibration VCMLE NO2
field calibration

SO2

FieldCalibration MLE SO2

Pandora115s1_FieldCalibration_MLE_SO2.jpeg - 105 KB
FieldCalibration MLE SO2

FieldCalibration VCMLE SO2

Pandora115s1_FieldCalibration_VCMLE_SO2.jpeg - 102 KB
FieldCalibration VCMLE SO2

Automatically harvested metadata

Automatically Harvested Metadata

Manual review and decision

Manual Review and Decision

Calibration notes, data-quality findings, and whole-period evaluation.

Data Evaluation

Could there be an offset between VPs?

No

Are there breakpoint-analysis plots or suspicious discontinuities?

No

Do daily figures before and after the new VP look consistent?

No

Is there an AMF dependency for one or more VPs?

No

Is there seasonality, for example negative values in winter?

No

Does the daily shape follow the trop data roughly?

No

If there is a discrepancy, what are the DS and Sky viewing directions?

No

Is there a co-located or nearby instrument for comparison?

No

O3Ext

Decision for O3Ext
With Care
most day look good, some show wiggly bahaviour

O3Synt

Decision for O3Synt
With Care

NO2

Decision for NO2
Yes
looks good, E-MLE used

SO2

Decision for SO2
Unusable
to high

HCHO

Decision for HCHO
Unusable
no HCHO cal done for this VP because of DELRIN in the headsensor

H2O

Decision for H2O
Yes
ok