Decision Summary

Upload-ready calibration protocol overview
Instrument P115s1
Session 27
Location LabIBK
Calibration person Hanna Rohringer
CF file Pandora115s1_CF_v27d20170313.txt
Software version 1.8.89
Calibration vector calV9
Reference day 20170806
Overall status development
O3Ext: With CareO3Synt: With CareNO2: YesSO2: With CareHCHO: UnusableH2O: Yes
Instrument P115s1
Session 27
Events 431
Figures 15
Action time 1h 5m

Action Timeline

73 action intervals
Feb 21, 2026Feb 23, 202604:04 PM10:08 AMCreateReferenceFieldCalibrationWavelengthCalibrationNDFiltersSensitivityCalibrationConvolutionAssignCalFileCreateReference 2s abortedFieldCalibration 3s abortedCreateReference 7s abortedWavelengthCalibration 13s finishedNDFilters 1m 16s abortedNDFilters 2m 12s abortedNDFilters 12m 43s finishedSensitivityCalibration 6m 55s finishedSensitivityCalibration 0s finishedSensitivityCalibration 0s finishedConvolution 59s finishedConvolution 0s finishedConvolution 0s finishedConvolution 0s finishedConvolution 0s finishedCreateReference 14m 49s finishedCreateReference 0s abortedCreateReference 23s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 1m 17s finishedCreateReference 0s abortedCreateReference 1m 54s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 3m 5s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 1m 6s abortedCreateReference 35s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 33s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 13s abortedCreateReference 10s abortedCreateReference 7s finishedCreateReference 0s finishedFieldCalibration 1m 19s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedCreateReference 6s abortedCreateReference 2s abortedFieldCalibration 0s staleFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 10s abortedFieldCalibration 1m 12s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s abortedFieldCalibration 13m 16s finishedFieldCalibration 0s abortedFieldCalibration 8s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedAssignCalFile 27s finished

Duration by Step

CreateReference24m 30sNDFilters16m 10sFieldCalibration16m 7sSensitivityCalibration6m 55sConvolution59sAssignCalFile27sWavelengthCalibration13s

Action Summary

Action Events Runs Aborted Time
AssignCalFile 3 1 0 27s
Convolution 16 5 0 59s
CreateReference 62 31 14 24m 30s
FieldCalibration 172 28 4 16m 7s
NDFilters 88 3 2 16m 10s
No action 11 0 0 0s
SensitivityCalibration 76 3 0 6m 55s
WavelengthCalibration 3 1 0 13s

Session Figures

15 figures from Google Drive/figures
calprotocol section

Wavelength Calibration

Slit function parameters

Pandora115s1_WavelengthCalibration_SlitFuncParam.jpeg - 159 KB
Slit function parameters
reference period

Reference Day Decision Support

Upload the FOV and L0 raw-count evidence used to justify the selected reference day before reviewing the CreateReference figures.

calprotocol section

Reference Period and CreateReference

Create reference L1CHARFILT

Pandora115s1_CreateReference_L1CHARFILT_OPEN_U340.jpeg - 164 KB
Create reference L1CHARFILT

Create reference SELSPECS

Pandora115s1_CreateReference_SELSPECS_OPEN_U340.jpeg - 136 KB
Create reference SELSPECS

Create reference SLERR (STRAY)

Pandora115s1_CreateReference_SLERR(STRAY)_OPEN_U340.jpeg - 119 KB
Create reference SLERR (STRAY)

Create reference AODEST

Pandora115s1_CreateReference_AODEST.jpeg - 65 KB
Create reference AODEST

Create reference SENSCORRFIN (SENS)

Pandora115s1_CreateReference_SENSCORRFIN(SENS)_OPEN_U340.jpeg - 124 KB
Create reference SENSCORRFIN (SENS)

Create reference DISPFIT (DISP)

Pandora115s1_CreateReference_DISPFIT(DISP).jpeg - 115 KB
Create reference DISPFIT (DISP)

Create reference WLCHANGE (DISP)

Pandora115s1_CreateReference_WLCHANGE(DISP)_OPEN_U340.jpeg - 101 KB
Create reference WLCHANGE (DISP)

Create reference SHIFT (CLOS)

Pandora115s1_CreateReference_SHIFT(CLOS)_OPEN_U340.jpeg - 121 KB
Create reference SHIFT (CLOS)
field calibration

O3

FieldCalibration VC O3

Pandora115s1_FieldCalibration_VC_O3.jpeg - 110 KB
FieldCalibration VC O3

FieldCalibration SCREFINT O3

Pandora115s1_FieldCalibration_SCREFINT_O3.jpeg - 73 KB
FieldCalibration SCREFINT O3
field calibration

NO2

NO2 E-MLE settings applied fw1out=(); fw2out=(); dqf2in=(); ddlim=(-1.000,-1.000); shilim(nm)=(-0.050,0.050); uscmax=1.825e-03; wrmsmax=2.260e+00; atvarmax(%)=2.000e+01; amflim=(1.000,5.000); amflimvar=(0.000,1.000); perc=50.00; ngroup=10; timeaver(min)=0.00; redu=2; scref=(1.775e-04,1.554e-04,1.590e-04); uscref=(2.599e-05,8.079e-06,1.190e-05)

FieldCalibration MLE NO2

Pandora115s1_FieldCalibration_MLE_NO2.jpeg - 170 KB
FieldCalibration MLE NO2

FieldCalibration VCMLE NO2

Pandora115s1_FieldCalibration_VCMLE_NO2.jpeg - 143 KB
FieldCalibration VCMLE NO2
field calibration

SO2

SO2 MLE settings applied fw1out=(); fw2out=(); dqf2in=(); ddlim=(-1.000,-1.000); shilim(nm)=(-0.050,0.050); uscmax=2.333e-02; wrmsmax=4.915e+00; atvarmax(%)=2.000e+01; amflim=(1.000,2.300); amflimvar=(0.000,1.000); perc=0.00; ngroup=10; timeaver(min)=0.00; redu=0; scref=(1.027e-04); uscref=(2.210e-05)

FieldCalibration MLE SO2

Pandora115s1_FieldCalibration_MLE_SO2.jpeg - 130 KB
FieldCalibration MLE SO2

FieldCalibration VCMLE SO2

Pandora115s1_FieldCalibration_VCMLE_SO2.jpeg - 141 KB
FieldCalibration VCMLE SO2

Automatically harvested metadata

Automatically Harvested Metadata

Manual review and decision

Manual Review and Decision

Calibration notes, data-quality findings, and whole-period evaluation.

Data Evaluation

Could there be an offset between VPs?

No

Are there breakpoint-analysis plots or suspicious discontinuities?

No

Do daily figures before and after the new VP look consistent?

No

Is there an AMF dependency for one or more VPs?

No

Is there seasonality, for example negative values in winter?

No

Does the daily shape follow the trop data roughly?

No

If there is a discrepancy, what are the DS and Sky viewing directions?

No

Is there a co-located or nearby instrument for comparison?

No

O3Ext

Decision for O3Ext
With Care
most day look good, some show wiggly bahaviour

O3Synt

Decision for O3Synt
With Care

NO2

Decision for NO2
Yes
looks good, E-MLE used

SO2

Decision for SO2
With Care
there is a data jump but its lower the 1e-4 thats fine

HCHO

Decision for HCHO
Unusable
no HCHO cal done for this VP because of DELRIN in the headsensor

H2O

Decision for H2O
Yes
ok