Decision Summary
Upload-ready calibration protocol overview
Instrument
P115s1
Session
26
Location
Rome-ISAC
Calibration person
Hanna Rohringer
CF file
Pandora115s1_CF_v26d20180925.txt
Software version
1.8.89
Calibration vector
calV7
Reference day
20200606
Overall status
development
O3Ext: With CareO3Synt: With CareNO2: UnusableSO2: With CareHCHO: UnusableH2O: Yes
Action Timeline
54 action intervalsDuration by Step
Action Summary
| Action | Events | Runs | Aborted | Time |
|---|---|---|---|---|
| AssignCalFile | 3 | 1 | 0 | 40s |
| CreateReference | 65 | 31 | 10 | 7m 25s |
| FieldCalibration | 137 | 21 | 1 | 1h 0m |
| No action | 12 | 0 | 0 | 0s |
Session Figures
15 figures from Google Drive/figures
calprotocol section
Wavelength Calibration
reference period
Reference Day Decision Support
Upload the FOV and L0 raw-count evidence used to justify the selected reference day before reviewing the CreateReference figures.
calprotocol section
Reference Period and CreateReference
Create reference L1CHARFILT
Pandora115s1_CreateReference_L1CHARFILT_OPEN_U340.jpeg - 146 KBCreate reference SELSPECS
Pandora115s1_CreateReference_SELSPECS_OPEN_U340.jpeg - 130 KBCreate reference SLERR (STRAY)
Pandora115s1_CreateReference_SLERR(STRAY)_OPEN_U340.jpeg - 127 KBCreate reference AODEST
Pandora115s1_CreateReference_AODEST.jpeg - 63 KBCreate reference SENSCORRFIN (SENS)
Pandora115s1_CreateReference_SENSCORRFIN(SENS)_OPEN_U340.jpeg - 110 KBCreate reference DISPFIT (DISP)
Pandora115s1_CreateReference_DISPFIT(DISP).jpeg - 100 KBCreate reference WLCHANGE (DISP)
Pandora115s1_CreateReference_WLCHANGE(DISP)_OPEN_U340.jpeg - 214 KBCreate reference SHIFT (CLOS)
Pandora115s1_CreateReference_SHIFT(CLOS)_OPEN_U340.jpeg - 176 KB
field calibration
O3
field calibration
NO2
field calibration
SO2
Automatically harvested metadata
Automatically Harvested Metadata
Manual review and decision
Manual Review and Decision
Calibration notes, data-quality findings, and whole-period evaluation.Data Evaluation
Could there be an offset between VPs?
No
Are there breakpoint-analysis plots or suspicious discontinuities?
No
Do daily figures before and after the new VP look consistent?
No
Is there an AMF dependency for one or more VPs?
No
Is there seasonality, for example negative values in winter?
No
Does the daily shape follow the trop data roughly?
No
If there is a discrepancy, what are the DS and Sky viewing directions?
No
Is there a co-located or nearby instrument for comparison?
No
O3Ext
Decision for O3Ext
With Care
wrong ND
O3Synt
Decision for O3Synt
With Care
NO2
Decision for NO2
Unusable
wrong ND
SO2
Decision for SO2
With Care
wrong ND
HCHO
Decision for HCHO
Unusable
no HCHO cal done for this VP because of DELRIN in the headsensor
H2O
Decision for H2O
Yes
ok