Decision Summary

Upload-ready calibration protocol overview
Instrument P115s1
Session 19
Location Rome-ISAC
Calibration person Hanna Rohringer
CF file Pandora115s1_CF_v18d20170316.txt
Software version 1.8.74
Calibration vector calV6
Reference day 20170621
Overall status development
O3Ext: With CareO3Synt: UnusableNO2: UnusableSO2: With CareHCHO: UnusableH2O: Unusable
Instrument P115s1
Session 19
Events 333
Figures 16
Action time 4h 31m

Action Timeline

88 action intervals
Feb 02, 2026Feb 03, 202610:33 AM12:46 PMTempDependenceWavelengthCalibrationCreateReferenceFieldCalibrationAssignCalFileTempDependence 2s abortedWavelengthCalibration 1h 42m finishedCreateReference 3m 0s abortedCreateReference 45s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 21s finishedCreateReference 0s abortedCreateReference 15s finishedCreateReference 12m 15s abortedCreateReference 20s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 2m 47s finishedCreateReference 0s abortedCreateReference 1m 46s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 15s finishedCreateReference 0s abortedCreateReference 13s finishedCreateReference 0s abortedCreateReference 6m 35s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 17s abortedCreateReference 2m 25s finishedCreateReference 0s abortedCreateReference 1m 18s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 20s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 10s finishedCreateReference 0s finishedCreateReference 3s abortedFieldCalibration 57s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s abortedFieldCalibration 2m 21s finishedFieldCalibration 0s abortedFieldCalibration 50s abortedFieldCalibration 29s abortedFieldCalibration 58s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s abortedFieldCalibration 2m 47s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 11m 30s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s abortedFieldCalibration 1h 32m finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 13s abortedFieldCalibration 6m 26s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 24s abortedFieldCalibration 5m 30s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s abortedFieldCalibration 10m 39s abortedAssignCalFile 49s finished

Duration by Step

FieldCalibration2h 15mWavelengthCalibration1h 42mCreateReference33m 6sAssignCalFile49sTempDependence2s

Action Summary

Action Events Runs Aborted Time
AssignCalFile 3 1 0 49s
CreateReference 79 39 15 33m 6s
FieldCalibration 230 46 10 2h 15m
No action 16 0 0 0s
TempDependence 2 1 1 2s
WavelengthCalibration 3 1 0 1h 42m

Session Figures

16 figures from Google Drive/figures
calprotocol section

Wavelength Calibration

Slit function parameters

Pandora115s1_WavelengthCalibration_SlitFuncParam.jpeg - 170 KB
Slit function parameters
reference period

Reference Day Decision Support

Upload the FOV and L0 raw-count evidence used to justify the selected reference day before reviewing the CreateReference figures.

calprotocol section

Reference Period and CreateReference

Create reference L1CHARFILT

Pandora115s1_CreateReference_L1CHARFILT_OPEN_U340.jpeg - 122 KB
Create reference L1CHARFILT

Create reference SELSPECS

Pandora115s1_CreateReference_SELSPECS_OPEN_U340.jpeg - 113 KB
Create reference SELSPECS

Create reference SLERR (STRAY)

Pandora115s1_CreateReference_SLERR(STRAY)_OPEN_U340.jpeg - 147 KB
Create reference SLERR (STRAY)

Create reference AODEST

Pandora115s1_CreateReference_AODEST.jpeg - 53 KB
Create reference AODEST

Create reference SENSCORRFIN (SENS)

Pandora115s1_CreateReference_SENSCORRFIN(SENS)_OPEN_U340.jpeg - 93 KB
Create reference SENSCORRFIN (SENS)

Create reference DISPFIT (DISP)

Pandora115s1_CreateReference_DISPFIT(DISP).jpeg - 148 KB
Create reference DISPFIT (DISP)

Create reference WLCHANGE (DISP)

Pandora115s1_CreateReference_WLCHANGE(DISP)_OPEN_U340.jpeg - 179 KB
Create reference WLCHANGE (DISP)

Create reference SHIFT (CLOS)

Pandora115s1_CreateReference_SHIFT(CLOS)_OPEN_U340.jpeg - 180 KB
Create reference SHIFT (CLOS)
field calibration

O3

FieldCalibration VC O3

Pandora115s1_FieldCalibration_VC_O3.jpeg - 159 KB
FieldCalibration VC O3

FieldCalibration SCREFINT O3

Pandora115s1_FieldCalibration_SCREFINT_O3.jpeg - 112 KB
FieldCalibration SCREFINT O3
field calibration

NO2

NO2 E-MLE settings applied fw1out=(); fw2out=(); dqf2in=(); ddlim=(-1.000,-1.000); shilim(nm)=(-0.050,0.050); uscmax=2.111e-03; wrmsmax=2.041e+00; atvarmax(%)=2.000e+01; amflim=(1.000,5.000); amflimvar=(0.000,1.000); perc=50.00; ngroup=20; timeaver(min)=0.00; redu=2; scref=(2.778e-04,3.239e-04,3.027e-04); uscref=(2.131e-05,1.621e-05,1.720e-05)

FieldCalibration MLE NO2

Pandora115s1_FieldCalibration_MLE_NO2.jpeg - 197 KB
FieldCalibration MLE NO2

FieldCalibration VCMLE NO2

Pandora115s1_FieldCalibration_VCMLE_NO2.jpeg - 160 KB
FieldCalibration VCMLE NO2
field calibration

SO2

SO2 MLE settings applied fw1out=(); fw2out=(); dqf2in=(); ddlim=(-1.000,-1.000); shilim(nm)=(-0.050,0.050); uscmax=1.979e-02; wrmsmax=6.036e-01; atvarmax(%)=2.000e+01; amflim=(1.000,2.500); amflimvar=(0.000,1.000); perc=0.00; ngroup=15; timeaver(min)=0.00; redu=0; scref=(8.022e-05); uscref=(2.559e-05)

FieldCalibration MLE SO2

Pandora115s1_FieldCalibration_MLE_SO2.jpeg - 138 KB
FieldCalibration MLE SO2

FieldCalibration VCMLE SO2

Pandora115s1_FieldCalibration_VCMLE_SO2.jpeg - 133 KB
FieldCalibration VCMLE SO2
field calibration

HCHO

FieldCalibration MLE HCHO

Pandora115s1_FieldCalibration_MLE_HCHO.jpeg - 91 KB
FieldCalibration MLE HCHO

Automatically harvested metadata

Automatically Harvested Metadata

Manual review and decision

Manual Review and Decision

Calibration notes, data-quality findings, and whole-period evaluation.

Data Evaluation

Could there be an offset between VPs?

No

Are there breakpoint-analysis plots or suspicious discontinuities?

No

Do daily figures before and after the new VP look consistent?

No

Is there an AMF dependency for one or more VPs?

No

Is there seasonality, for example negative values in winter?

No

Does the daily shape follow the trop data roughly?

No

If there is a discrepancy, what are the DS and Sky viewing directions?

No

Is there a co-located or nearby instrument for comparison?

No

O3Ext

Decision for O3Ext
With Care
extrapolation needed, strong AMF dependent shape

O3Synt

Decision for O3Synt
Unusable

NO2

Decision for NO2
Unusable
ok, MLE used; with E-MLE the values are higher; perc=0; ngroup = filtered data points /100. Unusable lab cal

SO2

Decision for SO2
With Care
wrong lab cal

HCHO

Decision for HCHO
Unusable
no HCHO cal done for this VP because of DELRIN in the headsensor

H2O

Decision for H2O
Unusable