Decision Summary

Upload-ready calibration protocol overview
Instrument P115s1
Session 17
Location -
Calibration person Hanna Rohringer
CF file Pandora115s1_CF_v17d20170316.txt
Software version 1.8.74
Calibration vector calV5
Reference day -
Overall status development
O3Ext: With CareO3Synt: UnusableNO2: YesSO2: UnusableHCHO: UnusableH2O: Yes
Instrument P115s1
Session 17
Events 24
Figures 16
Action time 2h 4m

Action Timeline

5 action intervals
Jan 25, 2026Jan 29, 202607:25 PM03:39 PMFieldCalibrationAssignCalFileFieldCalibration 9m 13s finishedFieldCalibration 0s abortedFieldCalibration 1m 16s abortedAssignCalFile 40s finishedFieldCalibration 1h 53m aborted

Duration by Step

FieldCalibration2h 3mAssignCalFile40s

Action Summary

Action Events Runs Aborted Time
AssignCalFile 3 1 0 40s
FieldCalibration 8 4 3 2h 3m
No action 13 0 0 0s

Session Figures

16 figures from Google Drive/figures
calprotocol section

Wavelength Calibration

Slit function parameters

Pandora115s1_WavelengthCalibration_SlitFuncParam.jpeg - 98 KB
Slit function parameters
reference period

Reference Day Decision Support

Upload the FOV and L0 raw-count evidence used to justify the selected reference day before reviewing the CreateReference figures.

calprotocol section

Reference Period and CreateReference

Create reference L1CHARFILT

Pandora115s1_CreateReference_L1CHARFILT_OPEN_U340.jpeg - 84 KB
Create reference L1CHARFILT

Create reference SELSPECS

Pandora115s1_CreateReference_SELSPECS_OPEN_U340.jpeg - 72 KB
Create reference SELSPECS

Create reference SLERR (STRAY)

Pandora115s1_CreateReference_SLERR(STRAY)_OPEN_U340.jpeg - 62 KB
Create reference SLERR (STRAY)

Create reference AODEST

Pandora115s1_CreateReference_AODEST.jpeg - 27 KB
Create reference AODEST

Create reference SENSCORRFIN (SENS)

Pandora115s1_CreateReference_SENSCORRFIN(SENS)_OPEN_U340.jpeg - 69 KB
Create reference SENSCORRFIN (SENS)

Create reference DISPFIT (DISP)

Pandora115s1_CreateReference_DISPFIT(DISP).jpeg - 62 KB
Create reference DISPFIT (DISP)

Create reference WLCHANGE (DISP)

Pandora115s1_CreateReference_WLCHANGE(DISP)_OPEN_U340.jpeg - 59 KB
Create reference WLCHANGE (DISP)

Create reference SHIFT (CLOS)

Pandora115s1_CreateReference_SHIFT(CLOS)_OPEN_U340.jpeg - 64 KB
Create reference SHIFT (CLOS)
field calibration

O3

FieldCalibration VC O3

Pandora115s1_FieldCalibration_VC_O3.jpeg - 61 KB
FieldCalibration VC O3

FieldCalibration SCREFINT O3

Pandora115s1_FieldCalibration_SCREFINT_O3.jpeg - 38 KB
FieldCalibration SCREFINT O3
field calibration

NO2

FieldCalibration MLE NO2

Pandora115s1_FieldCalibration_MLE_NO2.jpeg - 259 KB
FieldCalibration MLE NO2

FieldCalibration VCMLE NO2

Pandora115s1_FieldCalibration_VCMLE_NO2.jpeg - 71 KB
FieldCalibration VCMLE NO2
field calibration

SO2

SO2 MLE settings applied fw1out=(); fw2out=(); dqf2in=(); ddlim=(-1.000,-1.000); shilim(nm)=(-0.050,0.050); uscmax=2.095e-02; wrmsmax=4.114e+00; atvarmax(%)=2.000e+01; amflim=(1.000,3.500); amflimvar=(0.000,1.000); perc=50.00; ngroup=10; timeaver(min)=0.00; redu=0; scref=(4.598e-05); uscref=(4.055e-05)

FieldCalibration MLE SO2

Pandora115s1_FieldCalibration_MLE_SO2.jpeg - 98 KB
FieldCalibration MLE SO2

FieldCalibration VCMLE SO2

Pandora115s1_FieldCalibration_VCMLE_SO2.jpeg - 121 KB
FieldCalibration VCMLE SO2
field calibration

HCHO

FieldCalibration MLE HCHO

Pandora115s1_FieldCalibration_MLE_HCHO.jpeg - 42 KB
FieldCalibration MLE HCHO

Automatically harvested metadata

Automatically Harvested Metadata

Manual review and decision

Manual Review and Decision

Calibration notes, data-quality findings, and whole-period evaluation.

Data Evaluation

Could there be an offset between VPs?

No

Are there breakpoint-analysis plots or suspicious discontinuities?

No

Do daily figures before and after the new VP look consistent?

No

Is there an AMF dependency for one or more VPs?

No

Is there seasonality, for example negative values in winter?

No

Does the daily shape follow the trop data roughly?

No

If there is a discrepancy, what are the DS and Sky viewing directions?

No

Is there a co-located or nearby instrument for comparison?

No

O3Ext

Decision for O3Ext
With Care
extrapolation needed, relativley high AtmVar; very high O3 uncertainty: 4.3e-3; strong AMF dependent shape, looks like SL pb

O3Synt

Decision for O3Synt
Unusable

NO2

Decision for NO2
Yes
ok, MLE used; with E-MLE the values are higher; perc=0; ngroup = filtered data points /100

SO2

Decision for SO2
Unusable
not following SOPs when doing MLE to match the SL of the other period

HCHO

Decision for HCHO
Unusable
no HCHO cal done for this VP because of DELRIN in the headsensor

H2O

Decision for H2O
Yes