Decision Summary

Upload-ready calibration protocol overview
Instrument P114s1
Session 5
Location BuenosAires
Calibration person Manuel Gebetsberger
CF file Pandora114s1_CF_v1d20190919.txt
Software version 1.8.49
Calibration vector -
Reference day -
Overall status development
O3Ext: YesO3Synt: UnusableNO2: YesSO2: With CareHCHO: UnusableH2O: Unusable
Instrument P114s1
Session 5
Events 93
Figures 17
Action time 6m 5s

Action Timeline

15 action intervals
Apr 09, 2020Apr 10, 202008:01 PM07:24 AMCheckDispersionCreateReferenceAssignCalFileCheckDispersion 1m 13s finishedCheckDispersion 5s finishedCheckDispersion 0s finishedCheckDispersion 1m 24s finishedCreateReference 11s abortedCreateReference 2m 35s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedAssignCalFile 37s finished

Duration by Step

CreateReference2m 45sCheckDispersion2m 42sAssignCalFile37s

Action Summary

Action Events Runs Aborted Time
AssignCalFile 4 1 0 37s
CheckDispersion 14 4 0 2m 42s
CreateReference 67 10 2 2m 45s
No action 8 0 0 0s

Session Figures

17 figures from Google Drive/figures
calprotocol section

Wavelength Calibration

Slit function parameters

Pandora114s1_WavelengthCalibration_SlitFuncParam.jpeg - 164 KB
Slit function parameters
reference period

Reference Day Decision Support

Upload the FOV and L0 raw-count evidence used to justify the selected reference day before reviewing the CreateReference figures.

calprotocol section

Reference Period and CreateReference

Create reference L1CHARFILT

Pandora114s1_CreateReference_L1CHARFILT_OPEN_U340.jpeg - 143 KB
Create reference L1CHARFILT

Create reference SELSPECS

Pandora114s1_CreateReference_SELSPECS_OPEN_U340.jpeg - 125 KB
Create reference SELSPECS

Create reference SLERR (STRAY)

Pandora114s1_CreateReference_SLERR(STRAY)_OPEN_U340.jpeg - 127 KB
Create reference SLERR (STRAY)

Create reference AODEST

Pandora114s1_CreateReference_AODEST.jpeg - 53 KB
Create reference AODEST

Create reference SENSCORRFIN (SENS)

Pandora114s1_CreateReference_SENSCORRFIN(SENS)_OPEN_U340.jpeg - 116 KB
Create reference SENSCORRFIN (SENS)

Create reference DISPFIT (DISP)

Pandora114s1_CreateReference_DISPFIT(DISP).jpeg - 113 KB
Create reference DISPFIT (DISP)

Create reference WLCHANGE (DISP)

Pandora114s1_CreateReference_WLCHANGE(DISP)_OPEN_U340.jpeg - 203 KB
Create reference WLCHANGE (DISP)

Create reference SHIFT (CLOS)

Pandora114s1_CreateReference_SHIFT(CLOS)_OPEN_U340.jpeg - 118 KB
Create reference SHIFT (CLOS)
field calibration

O3

FieldCalibration VC O3

Pandora114s1_FieldCalibration_VC_O3.jpeg - 103 KB
FieldCalibration VC O3

FieldCalibration SCREFINT O3

Pandora114s1_FieldCalibration_SCREFINT_O3.jpeg - 76 KB
FieldCalibration SCREFINT O3
field calibration

NO2

FieldCalibration MLE NO2

Pandora114s1_FieldCalibration_MLE_NO2.jpeg - 147 KB
FieldCalibration MLE NO2

FieldCalibration VCMLE NO2

Pandora114s1_FieldCalibration_VCMLE_NO2.jpeg - 99 KB
FieldCalibration VCMLE NO2
field calibration

SO2

FieldCalibration MLE SO2

Pandora114s1_FieldCalibration_MLE_SO2.jpeg - 108 KB
FieldCalibration MLE SO2

FieldCalibration VCMLE SO2

Pandora114s1_FieldCalibration_VCMLE_SO2.jpeg - 89 KB
FieldCalibration VCMLE SO2
field calibration

HCHO

FieldCalibration MLE HCHO

Pandora114s1_FieldCalibration_MLE_HCHO.jpeg - 105 KB
FieldCalibration MLE HCHO

FieldCalibration VCMLE HCHO

Pandora114s1_FieldCalibration_VCMLE_HCHO.jpeg - 83 KB
FieldCalibration VCMLE HCHO

Automatically harvested metadata

Automatically Harvested Metadata

Manual review and decision

Manual Review and Decision

Calibration notes, data-quality findings, and whole-period evaluation.

Data Evaluation

Could there be an offset between VPs?

No

Are there breakpoint-analysis plots or suspicious discontinuities?

No

Do daily figures before and after the new VP look consistent?

No

Is there an AMF dependency for one or more VPs?

No

Is there seasonality, for example negative values in winter?

No

Does the daily shape follow the trop data roughly?

No

If there is a discrepancy, what are the DS and Sky viewing directions?

No

Is there a co-located or nearby instrument for comparison?

No

O3Ext

Decision for O3Ext
Yes
O3Ext noise but usable - O3Synt and O3T are unusable

O3Synt

Decision for O3Synt
Unusable

NO2

Decision for NO2
Yes
Needs another validity period

SO2

Decision for SO2
With Care

HCHO

Decision for HCHO
Unusable

H2O

Decision for H2O
Unusable