Decision Summary
Upload-ready calibration protocol overview
Instrument
P114s1
Session
3
Location
BuenosAires
Calibration person
Manuel Gebetsberger
CF file
Pandora114s1_CF_v3d20170314.txt
Software version
1.8.26
Calibration vector
-
Reference day
-
Overall status
development
O3Ext: YesO3Synt: UnusableNO2: With CareSO2: With CareHCHO: UnusableH2O: With Care
Action Timeline
26 action intervalsDuration by Step
Action Summary
| Action | Events | Runs | Aborted | Time |
|---|---|---|---|---|
| AssignCalFile | 3 | 1 | 0 | 30s |
| CheckDispersion | 16 | 4 | 0 | 5m 3s |
| CreateReference | 95 | 21 | 10 | 1h 49m |
| No action | 13 | 0 | 0 | 0s |
Session Figures
1 figure from Google Drive/figuresAutomatically harvested metadata
Automatically Harvested Metadata
Manual review and decision
Manual Review and Decision
Calibration notes, data-quality findings, and whole-period evaluation.Data Evaluation
Could there be an offset between VPs?
No
Are there breakpoint-analysis plots or suspicious discontinuities?
No
Do daily figures before and after the new VP look consistent?
No
Is there an AMF dependency for one or more VPs?
No
Is there seasonality, for example negative values in winter?
No
Does the daily shape follow the trop data roughly?
No
If there is a discrepancy, what are the DS and Sky viewing directions?
No
Is there a co-located or nearby instrument for comparison?
No
O3Ext
Decision for O3Ext
Yes
O3Synt is almost all flagged out. O3Ext looks ok.
O3Synt
Decision for O3Synt
Unusable
NO2
Decision for NO2
With Care
E-MLE, only half the period usable
SO2
Decision for SO2
With Care
Only half period usable
HCHO
Decision for HCHO
Unusable
All flagged out, delrin period
H2O
Decision for H2O
With Care
okeish