Decision Summary

Upload-ready calibration protocol overview
Instrument P114s1
Session 14
Location -
Calibration person SMorhenn
CF file -
Software version 1.8.66
Calibration vector -
Reference day -
Overall status development
O3Ext: YesO3Synt: With CareNO2: YesSO2: YesHCHO: UnusableH2O: Not answered
Instrument P114s1
Session 14
Events 0
Figures 16
Action time 0s

Action Timeline

No start/finish intervals found.
No timeline intervals found

Duration by Step

No completed durations found

Action Summary

Action Events Runs Aborted Time
No actions loaded

Session Figures

16 figures from Google Drive/figures
calprotocol section

Wavelength Calibration

Slit function parameters

Pandora114s1_WavelengthCalibration_SlitFuncParam.jpeg - 232 KB
Slit function parameters
reference period

Reference Day Decision Support

Upload the FOV and L0 raw-count evidence used to justify the selected reference day before reviewing the CreateReference figures.

calprotocol section

Reference Period and CreateReference

Create reference L1CHARFILT

Pandora114s1_CreateReference_L1CHARFILT_OPEN_U340.jpeg - 140 KB
Create reference L1CHARFILT

Create reference SELSPECS

Pandora114s1_CreateReference_SELSPECS_OPEN_U340.jpeg - 128 KB
Create reference SELSPECS

Create reference SLERR (STRAY)

Pandora114s1_CreateReference_SLERR(STRAY)_OPEN_U340.jpeg - 127 KB
Create reference SLERR (STRAY)

Create reference AODEST

Pandora114s1_CreateReference_AODEST.jpeg - 69 KB
Create reference AODEST

Create reference SENSCORRFIN (SENS)

Pandora114s1_CreateReference_SENSCORRFIN(SENS)_OPEN_U340.jpeg - 124 KB
Create reference SENSCORRFIN (SENS)

Create reference DISPFIT (DISP)

Pandora114s1_CreateReference_DISPFIT(DISP).jpeg - 174 KB
Create reference DISPFIT (DISP)

Create reference WLCHANGE (DISP)

Pandora114s1_CreateReference_WLCHANGE(DISP)_OPEN_U340.jpeg - 169 KB
Create reference WLCHANGE (DISP)

Create reference SHIFT (CLOS)

Pandora114s1_CreateReference_SHIFT(CLOS)_OPEN_U340.jpeg - 191 KB
Create reference SHIFT (CLOS)
field calibration

O3

FieldCalibration VC O3

Pandora114s1_FieldCalibration_VC_O3.jpeg - 166 KB
FieldCalibration VC O3

FieldCalibration SCREFINT O3

Pandora114s1_FieldCalibration_SCREFINT_O3.jpeg - 111 KB
FieldCalibration SCREFINT O3
field calibration

NO2

FieldCalibration MLE NO2

Pandora114s1_FieldCalibration_MLE_NO2.jpeg - 233 KB
FieldCalibration MLE NO2

FieldCalibration VCMLE NO2

Pandora114s1_FieldCalibration_VCMLE_NO2.jpeg - 223 KB
FieldCalibration VCMLE NO2
field calibration

SO2

FieldCalibration MLE SO2

Pandora114s1_FieldCalibration_MLE_SO2.jpeg - 170 KB
FieldCalibration MLE SO2

FieldCalibration VCMLE SO2

Pandora114s1_FieldCalibration_VCMLE_SO2.jpeg - 150 KB
FieldCalibration VCMLE SO2
field calibration

HCHO

FieldCalibration MLE HCHO

Pandora114s1_FieldCalibration_MLE_HCHO.jpeg - 86 KB
FieldCalibration MLE HCHO

Automatically harvested metadata

Automatically Harvested Metadata

Manual review and decision

Manual Review and Decision

Calibration notes, data-quality findings, and whole-period evaluation.

Data Evaluation

Could there be an offset between VPs?

No

Are there breakpoint-analysis plots or suspicious discontinuities?

No

Do daily figures before and after the new VP look consistent?

No

Is there an AMF dependency for one or more VPs?

No

Is there seasonality, for example negative values in winter?

No

Does the daily shape follow the trop data roughly?

No

If there is a discrepancy, what are the DS and Sky viewing directions?

No

Is there a co-located or nearby instrument for comparison?

No

O3Ext

Decision for O3Ext
Yes
O3Temp too low (almost 10K lower than in session13), extrapolation used in O3 cal

O3Synt

Decision for O3Synt
With Care

NO2

Decision for NO2
Yes
E-MLE used

SO2

Decision for SO2
Yes
looks good

HCHO

Decision for HCHO
Unusable
Delrin instrument

H2O

Decision for H2O