Decision Summary
Upload-ready calibration protocol overview
Instrument
P111s1
Session
3
Location
-
Calibration person
Moritz Mueller
CF file
C:/Blick/data/calresults/Pandora111s1/session3/Pandora111s1_CF_session3.txt
Software version
1.8.26
Calibration vector
-
Reference day
-
Overall status
development
O3Ext: With CareO3Synt: With CareNO2: YesSO2: With CareHCHO: UnusableH2O: Not answered
Action Timeline
No start/finish intervals found.No timeline intervals found
Duration by Step
No completed durations found
Action Summary
| Action | Events | Runs | Aborted | Time |
|---|---|---|---|---|
| No action | 7 | 0 | 0 | 0s |
Session Figures
15 figures from Google Drive/figures
calprotocol section
Wavelength Calibration
reference period
Reference Day Decision Support
Upload the FOV and L0 raw-count evidence used to justify the selected reference day before reviewing the CreateReference figures.
calprotocol section
Reference Period and CreateReference
Create reference L1CHARFILT
Pandora111s1_CreateReference_L1CHARFILT_OPEN_U340.jpeg - 80 KBCreate reference SELSPECS
Pandora111s1_CreateReference_SELSPECS_OPEN_U340.jpeg - 63 KBCreate reference SLERR (STRAY)
Pandora111s1_CreateReference_SLERR(STRAY)_OPEN_U340.jpeg - 60 KBCreate reference AODEST
Pandora111s1_CreateReference_AODEST.jpeg - 27 KBCreate reference SENSCORRFIN (SENS)
Pandora111s1_CreateReference_SENSCORRFIN(SENS)_OPEN_U340.jpeg - 87 KBCreate reference DISPFIT (DISP)
Pandora111s1_CreateReference_DISPFIT(DISP).jpeg - 60 KBCreate reference WLCHANGE (DISP)
Pandora111s1_CreateReference_WLCHANGE(DISP)_OPEN_U340.jpeg - 64 KBCreate reference SHIFT (CLOS)
Pandora111s1_CreateReference_SHIFT(CLOS)_OPEN_U340.jpeg - 78 KB
field calibration
O3
field calibration
NO2
field calibration
SO2
Automatically harvested metadata
Automatically Harvested Metadata
Manual review and decision
Manual Review and Decision
Calibration notes, data-quality findings, and whole-period evaluation.Data Evaluation
Could there be an offset between VPs?
No
Are there breakpoint-analysis plots or suspicious discontinuities?
No
Do daily figures before and after the new VP look consistent?
No
Is there an AMF dependency for one or more VPs?
No
Is there seasonality, for example negative values in winter?
No
Does the daily shape follow the trop data roughly?
No
If there is a discrepancy, what are the DS and Sky viewing directions?
No
Is there a co-located or nearby instrument for comparison?
No
O3Ext
Decision for O3Ext
With Care
was difficult to find a good reference day, 1 July 2019 was chosen (low pollution day)
O3Synt
Decision for O3Synt
With Care
NO2
Decision for NO2
Yes
EMLE used. Although the MLE (including uncertainty) looked better than the EMLE, the NO2 product was clearly better and more consistent with sky data when using the EMLE.
SO2
Decision for SO2
With Care
For the SO2 MLE, perc=10.00 gave a reasonable sref amount
HCHO
Decision for HCHO
Unusable
Delrin, not calibrated
H2O
Decision for H2O