Decision Summary

Upload-ready calibration protocol overview
Instrument P111s1
Session 10
Location INOE
Calibration person Moritz Mueller
CF file C:/Blick/data/calresults/Pandora111s1/session10/Pandora111s1_CF_session10.txt
Software version 1.8.58
Calibration vector -
Reference day -
Overall status development
O3Ext: With CareO3Synt: With CareNO2: YesSO2: With CareHCHO: UnusableH2O: Yes
Instrument P111s1
Session 10
Events 26
Figures 1
Action time 2m 4s

Action Timeline

7 action intervals
Dec 19, 201803:30 PM03:49 PMCheckDispersionCheckDispersion 9s abortedCheckDispersion 39s finishedCheckDispersion 5s finishedCheckDispersion 0s finishedCheckDispersion 41s abortedCheckDispersion 10s abortedCheckDispersion 21s finished

Duration by Step

CheckDispersion2m 4s

Action Summary

Action Events Runs Aborted Time
CheckDispersion 20 7 3 2m 4s
No action 6 0 0 0s

Session Figures

1 figure from Google Drive/figures
calprotocol section

Wavelength Calibration

Slit function parameters

Pandora111s1_WavelengthCalibration_SlitFuncParam(A2).jpeg - 54 KB
Slit function parameters

Automatically harvested metadata

Automatically Harvested Metadata

Manual review and decision

Manual Review and Decision

Calibration notes, data-quality findings, and whole-period evaluation.

Data Evaluation

Could there be an offset between VPs?

No

Are there breakpoint-analysis plots or suspicious discontinuities?

No

Do daily figures before and after the new VP look consistent?

No

Is there an AMF dependency for one or more VPs?

No

Is there seasonality, for example negative values in winter?

No

Does the daily shape follow the trop data roughly?

No

If there is a discrepancy, what are the DS and Sky viewing directions?

No

Is there a co-located or nearby instrument for comparison?

No

O3Ext

Decision for O3Ext
With Care
T too high wrt climatology - extRef: low Q - high wrms

O3Synt

Decision for O3Synt
With Care

NO2

Decision for NO2
Yes
not too easy to define limits: best guess based on MLE-EMLE-comparison

SO2

Decision for SO2
With Care
low SO2 - used perc 10 rather than 0 - certain days below 0

HCHO

Decision for HCHO
Unusable
Delrin, not calibrated

H2O

Decision for H2O
Yes
seems ok