Decision Summary
Upload-ready calibration protocol overview
Instrument
P109s1
Session
3
Location
Aldino
Calibration person
Lena Shalaby
CF file
Pandora109s1_CF_v3d20161108.txt
Software version
1.8.56
Calibration vector
-
Reference day
-
Overall status
development
O3Ext: Not answeredO3Synt: Not answeredNO2: Not answeredSO2: Not answeredHCHO: Not answeredH2O: Not answered
Instrument
P109s1
Session
3
Events
1478
Figures
1
Action time
8h 26m
Action Timeline
239 action intervals
Duration by Step
Action Summary
| Action |
Events |
Runs |
Aborted |
Time |
| AnalyzeBright |
77 |
10 |
6 |
1h 32m |
| AnalyzeDark |
98 |
21 |
19 |
28m 31s |
| AssignCalFile |
6 |
2 |
0 |
3m 39s |
| CheckDispersion |
27 |
8 |
0 |
4m 7s |
| CheckFilters |
18 |
6 |
3 |
3m 0s |
| CheckNoiseWave |
18 |
4 |
2 |
1m 38s |
| Convolution |
48 |
15 |
0 |
3m 22s |
| CreateReference |
97 |
10 |
1 |
14m 23s |
| NDFilters |
536 |
29 |
26 |
1h 36m |
| No action |
32 |
0 |
0 |
0s |
| SensitivityCalibration |
99 |
7 |
2 |
1h 14m |
| StrayLight |
3 |
1 |
1 |
3s |
| TempDependence |
24 |
5 |
3 |
3m 52s |
| WavelengthCalibration |
395 |
119 |
64 |
3h 1m |
Session Figures
1 figure from Google Drive/figures
Automatically harvested metadata
Automatically Harvested Metadata
Manual review and decision
Manual Review and Decision
Calibration notes, data-quality findings, and whole-period evaluation.
Data Evaluation
Could there be an offset between VPs?
No
Are there breakpoint-analysis plots or suspicious discontinuities?
No
Do daily figures before and after the new VP look consistent?
No
Is there an AMF dependency for one or more VPs?
No
Is there seasonality, for example negative values in winter?
No
Does the daily shape follow the trop data roughly?
No
If there is a discrepancy, what are the DS and Sky viewing directions?
No
Is there a co-located or nearby instrument for comparison?
No