Decision Summary
Upload-ready calibration protocol overview
Instrument
P106s1
Session
2
Location
-
Calibration person
RRajagopalan
CF file
-
Software version
1.8.29
Calibration vector
-
Reference day
-
Overall status
development
O3Ext: YesO3Synt: UnusableNO2: YesSO2: YesHCHO: With CareH2O: Not answered
Action Timeline
No start/finish intervals found.No timeline intervals found
Duration by Step
No completed durations found
Action Summary
| Action | Events | Runs | Aborted | Time |
|---|---|---|---|---|
| No actions loaded | ||||
Session Figures
1 figure from Google Drive/figuresAutomatically harvested metadata
Automatically Harvested Metadata
Manual review and decision
Manual Review and Decision
Calibration notes, data-quality findings, and whole-period evaluation.Data Evaluation
Could there be an offset between VPs?
No
Are there breakpoint-analysis plots or suspicious discontinuities?
No
Do daily figures before and after the new VP look consistent?
No
Is there an AMF dependency for one or more VPs?
No
Is there seasonality, for example negative values in winter?
No
Does the daily shape follow the trop data roughly?
No
If there is a discrepancy, what are the DS and Sky viewing directions?
No
Is there a co-located or nearby instrument for comparison?
No
O3Ext
Decision for O3Ext
Yes
Ushape in O3 temperature
O3Synt
Decision for O3Synt
Unusable
NO2
Decision for NO2
Yes
okay
SO2
Decision for SO2
Yes
okay
HCHO
Decision for HCHO
With Care
hard to compare without trop data, but look "reasonable"
H2O
Decision for H2O
removed it from calibration since there was no proper MLE possible and after the CF assignment the total columns looked invers-Ushape altough the final BlickC plot looked okay.