Decision Summary

Upload-ready calibration protocol overview
Instrument P106s1
Session 24
Location -
Calibration person LHaunold
CF file -
Software version 1.8.85
Calibration vector -
Reference day -
Overall status development
O3Ext: YesO3Synt: With CareNO2: YesSO2: With CareHCHO: UnusableH2O: Unusable
Instrument P106s1
Session 24
Events 0
Figures 9
Action time 0s

Action Timeline

No start/finish intervals found.
No timeline intervals found

Duration by Step

No completed durations found

Action Summary

Action Events Runs Aborted Time
No actions loaded

Session Figures

9 figures from Google Drive/figures
calprotocol section

Wavelength Calibration

Slit function parameters

Pandora106s1_WavelengthCalibration_SlitFuncParam.jpeg - 120 KB
Slit function parameters
reference period

Reference Day Decision Support

Upload the FOV and L0 raw-count evidence used to justify the selected reference day before reviewing the CreateReference figures.

calprotocol section

Reference Period and CreateReference

Create reference L1CHARFILT

Pandora106s1_CreateReference_L1CHARFILT_OPEN_U340.jpeg - 122 KB
Create reference L1CHARFILT

Create reference SELSPECS

Pandora106s1_CreateReference_SELSPECS_OPEN_U340.jpeg - 73 KB
Create reference SELSPECS

Create reference SLERR (STRAY)

Pandora106s1_CreateReference_SLERR(STRAY)_OPEN_U340.jpeg - 77 KB
Create reference SLERR (STRAY)

Create reference AODEST

Pandora106s1_CreateReference_AODEST.jpeg - 45 KB
Create reference AODEST

Create reference SENSCORRFIN (SENS)

Pandora106s1_CreateReference_SENSCORRFIN(SENS)_OPEN_U340.jpeg - 80 KB
Create reference SENSCORRFIN (SENS)

Create reference SHIFT (CLOS)

Pandora106s1_CreateReference_SHIFT(CLOS)_OPEN_U340.jpeg - 120 KB
Create reference SHIFT (CLOS)
field calibration

O3

FieldCalibration VC O3

Pandora106s1_FieldCalibration_VC_O3.jpeg - 71 KB
FieldCalibration VC O3

FieldCalibration SCREFINT O3

Pandora106s1_FieldCalibration_SCREFINT_O3.jpeg - 54 KB
FieldCalibration SCREFINT O3

Automatically harvested metadata

Automatically Harvested Metadata

Manual review and decision

Manual Review and Decision

Calibration notes, data-quality findings, and whole-period evaluation.

Data Evaluation

Could there be an offset between VPs?

No

Are there breakpoint-analysis plots or suspicious discontinuities?

No

Do daily figures before and after the new VP look consistent?

No

Is there an AMF dependency for one or more VPs?

No

Is there seasonality, for example negative values in winter?

No

Does the daily shape follow the trop data roughly?

No

If there is a discrepancy, what are the DS and Sky viewing directions?

No

Is there a co-located or nearby instrument for comparison?

No

O3Ext

Decision for O3Ext
Yes

O3Synt

Decision for O3Synt
With Care

NO2

Decision for NO2
Yes

SO2

Decision for SO2
With Care

HCHO

Decision for HCHO
Unusable
delrin

H2O

Decision for H2O
Unusable
very inverse-u shaped, tried different slit function characterizations but nothing improved