Decision Summary
Upload-ready calibration protocol overview
Instrument
P105s1
Session
6
Location
Helsinki
Calibration person
Markus Kilian
CF file
Pandora105s1_CF_v6d20160627.txt
Software version
1.8.75
Calibration vector
calV1
Reference day
20160908
Overall status
development
O3Ext: With CareO3Synt: With CareNO2: YesSO2: UnusableHCHO: UnusableH2O: Not answered
Action Timeline
65 action intervalsDuration by Step
Action Summary
| Action | Events | Runs | Aborted | Time |
|---|---|---|---|---|
| AssignCalFile | 6 | 2 | 0 | 1m 38s |
| Convolution | 30 | 10 | 0 | 3m 33s |
| CreateReference | 99 | 47 | 16 | 10m 44s |
| FieldCalibration | 5 | 3 | 1 | 32s |
| No action | 17 | 0 | 0 | 0s |
| SensitivityCalibration | 63 | 3 | 0 | 30s |
Session Figures
17 figures from Google Drive/figures
calprotocol section
Wavelength Calibration
reference period
Reference Day Decision Support
Upload the FOV and L0 raw-count evidence used to justify the selected reference day before reviewing the CreateReference figures.
calprotocol section
Reference Period and CreateReference
Create reference L1CHARFILT
Pandora105s1_CreateReference_L1CHARFILT_OPEN_U340.jpeg - 117 KBCreate reference SELSPECS
Pandora105s1_CreateReference_SELSPECS_OPEN_U340.jpeg - 87 KBCreate reference SLERR (STRAY)
Pandora105s1_CreateReference_SLERR(STRAY)_OPEN_U340.jpeg - 76 KBCreate reference AODEST
Pandora105s1_CreateReference_AODEST.jpeg - 47 KBCreate reference SENSCORRFIN (SENS)
Pandora105s1_CreateReference_SENSCORRFIN(SENS)_OPEN_U340.jpeg - 85 KBCreate reference DISPFIT (DISP)
Pandora105s1_CreateReference_DISPFIT(DISP).jpeg - 73 KBCreate reference WLCHANGE (DISP)
Pandora105s1_CreateReference_WLCHANGE(DISP)_OPEN_U340.jpeg - 83 KBCreate reference SHIFT (CLOS)
Pandora105s1_CreateReference_SHIFT(CLOS)_OPEN_U340.jpeg - 85 KB
field calibration
O3
field calibration
NO2
field calibration
SO2
field calibration
HCHO
Automatically harvested metadata
Automatically Harvested Metadata
Manual review and decision
Manual Review and Decision
Calibration notes, data-quality findings, and whole-period evaluation.Data Evaluation
Could there be an offset between VPs?
No
Are there breakpoint-analysis plots or suspicious discontinuities?
No
Do daily figures before and after the new VP look consistent?
No
Is there an AMF dependency for one or more VPs?
No
Is there seasonality, for example negative values in winter?
No
Does the daily shape follow the trop data roughly?
No
If there is a discrepancy, what are the DS and Sky viewing directions?
No
Is there a co-located or nearby instrument for comparison?
No
O3Ext
Decision for O3Ext
With Care
interpol needed, unct=3.9e-4
O3Synt
Decision for O3Synt
With Care
NO2
Decision for NO2
Yes
no sky data available
SO2
Decision for SO2
Unusable
clearly lower so2 as other VPs
HCHO
Decision for HCHO
Unusable
Delrin period
H2O
Decision for H2O