Decision Summary
Upload-ready calibration protocol overview
Instrument
P105s1
Session
105
Location
-
Calibration person
-
CF file
-
Software version
1.8.16
Calibration vector
-
Reference day
-
Overall status
development
O3Ext: With CareO3Synt: With CareNO2: YesSO2: With CareHCHO: YesH2O: Not answered
Action Timeline
No start/finish intervals found.No timeline intervals found
Duration by Step
No completed durations found
Action Summary
| Action | Events | Runs | Aborted | Time |
|---|---|---|---|---|
| No actions loaded | ||||
Session Figures
17 figures from Google Drive/figures
calprotocol section
Wavelength Calibration
reference period
Reference Day Decision Support
Upload the FOV and L0 raw-count evidence used to justify the selected reference day before reviewing the CreateReference figures.
calprotocol section
Reference Period and CreateReference
Create reference L1CHARFILT
Pandora105s1_CreateReference_L1CHARFILT_OPEN_U340.jpeg - 95 KBCreate reference SELSPECS
Pandora105s1_CreateReference_SELSPECS_OPEN_U340.jpeg - 86 KBCreate reference SLERR (STRAY)
Pandora105s1_CreateReference_SLERR(STRAY)_OPEN_U340.jpeg - 60 KBCreate reference AODEST
Pandora105s1_CreateReference_AODEST.jpeg - 27 KBCreate reference SENSCORRFIN (SENS)
Pandora105s1_CreateReference_SENSCORRFIN(SENS)_OPEN_U340.jpeg - 75 KBCreate reference DISPFIT (DISP)
Pandora105s1_CreateReference_DISPFIT(DISP).jpeg - 82 KBCreate reference WLCHANGE (DISP)
Pandora105s1_CreateReference_WLCHANGE(DISP)_OPEN_U340.jpeg - 75 KBCreate reference SHIFT (CLOS)
Pandora105s1_CreateReference_SHIFT(CLOS)_OPEN_U340.jpeg - 65 KB
field calibration
O3
field calibration
NO2
field calibration
SO2
field calibration
HCHO
Automatically harvested metadata
Automatically Harvested Metadata
Manual review and decision
Manual Review and Decision
Calibration notes, data-quality findings, and whole-period evaluation.Data Evaluation
Could there be an offset between VPs?
No
Are there breakpoint-analysis plots or suspicious discontinuities?
No
Do daily figures before and after the new VP look consistent?
No
Is there an AMF dependency for one or more VPs?
No
Is there seasonality, for example negative values in winter?
No
Does the daily shape follow the trop data roughly?
No
If there is a discrepancy, what are the DS and Sky viewing directions?
No
Is there a co-located or nearby instrument for comparison?
No
O3Ext
Decision for O3Ext
With Care
unct=2e-4; no interpol needed; out strongly SL affected (as we remember from 1.7 data which already had strong SL) so I included the SL calibration; the wrms for out for this VP is higher than for the other VPs, I do not know what the reason is maybe it is related to the slit function gotten from the spectral lamp measurements since the spec had been opened at some point
O3Synt
Decision for O3Synt
With Care
NO2
Decision for NO2
Yes
ok
SO2
Decision for SO2
With Care
ok, might be SL affected due to high SL of the system; seems to be higher as other VPs, SL?
HCHO
Decision for HCHO
Yes
ok
H2O
Decision for H2O