Decision Summary

Upload-ready calibration protocol overview
Instrument P104s1
Session 8
Location Downsview
Calibration person Ilias Bougoudis
CF file Pandora104s1_CF_v8d20221101.txt
Software version p1-8
Calibration vector -
Reference day 20230529
Overall status development
O3Ext: With CareO3Synt: With CareNO2: With CareSO2: With CareHCHO: With CareH2O: Yes
Instrument P104s1
Session 8
Events 275
Figures 17
Action time 1h 59m

Action Timeline

60 action intervals
Jun 26, 202301:02 PM03:53 PMCreateReferenceFieldCalibrationAssignCalFileCreateReference 33s finishedCreateReference 0s abortedCreateReference 32s finishedCreateReference 0s abortedCreateReference 28s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 20s finishedCreateReference 0s abortedCreateReference 1m 19s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 1m 42s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 52s abortedCreateReference 19s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 20s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 56s finishedCreateReference 0s finishedFieldCalibration 2m 13s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 1h 33m finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 8m 11s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 8m 19s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedAssignCalFile 43s finished

Duration by Step

FieldCalibration1h 51mCreateReference7m 21sAssignCalFile43s

Action Summary

Action Events Runs Aborted Time
AssignCalFile 3 1 0 43s
CreateReference 53 27 9 7m 21s
FieldCalibration 211 32 0 1h 51m
No action 8 0 0 0s

Session Figures

17 figures from Google Drive/figures
calprotocol section

Wavelength Calibration

Slit function parameters

Pandora104s1_WavelengthCalibration_SlitFuncParam.jpeg - 168 KB
Slit function parameters
reference period

Reference Day Decision Support

Upload the FOV and L0 raw-count evidence used to justify the selected reference day before reviewing the CreateReference figures.

calprotocol section

Reference Period and CreateReference

Create reference L1CHARFILT

Pandora104s1_CreateReference_L1CHARFILT_OPEN_U340.jpeg - 145 KB
Create reference L1CHARFILT

Create reference SELSPECS

Pandora104s1_CreateReference_SELSPECS_OPEN_U340.jpeg - 117 KB
Create reference SELSPECS

Create reference SLERR (STRAY)

Pandora104s1_CreateReference_SLERR(STRAY)_OPEN_U340.jpeg - 126 KB
Create reference SLERR (STRAY)

Create reference AODEST

Pandora104s1_CreateReference_AODEST.jpeg - 54 KB
Create reference AODEST

Create reference SENSCORRFIN (SENS)

Pandora104s1_CreateReference_SENSCORRFIN(SENS)_OPEN_U340.jpeg - 120 KB
Create reference SENSCORRFIN (SENS)

Create reference DISPFIT (DISP)

Pandora104s1_CreateReference_DISPFIT(DISP).jpeg - 108 KB
Create reference DISPFIT (DISP)

Create reference WLCHANGE (DISP)

Pandora104s1_CreateReference_WLCHANGE(DISP)_OPEN_U340.jpeg - 116 KB
Create reference WLCHANGE (DISP)

Create reference SHIFT (CLOS)

Pandora104s1_CreateReference_SHIFT(CLOS)_OPEN_U340.jpeg - 115 KB
Create reference SHIFT (CLOS)
field calibration

O3

FieldCalibration VC O3

Pandora104s1_FieldCalibration_VC_O3.jpeg - 120 KB
FieldCalibration VC O3

FieldCalibration SCREFINT O3

Pandora104s1_FieldCalibration_SCREFINT_O3.jpeg - 76 KB
FieldCalibration SCREFINT O3
field calibration

NO2

NO2 E-MLE settings applied fw1out=(); fw2out=(); dqf2in=(); ddlim=(-1.000,-1.000); shilim(nm)=(-0.050,0.050); uscmax=1.049e-03; wrmsmax=9.000e-04; atvarmax(%)=2.000e+01; amflim=(1.000,5.000); amflimvar=(0.000,1.000); perc=50.00; ngroup=71; timeaver(min)=0.00; redu=2; scref=(1.756e-04,1.614e-04,1.706e-04); uscref=(8.866e-06,1.041e-05,7.905e-06)

FieldCalibration MLE NO2

Pandora104s1_FieldCalibration_MLE_NO2.jpeg - 175 KB
FieldCalibration MLE NO2

FieldCalibration VCMLE NO2

Pandora104s1_FieldCalibration_VCMLE_NO2.jpeg - 108 KB
FieldCalibration VCMLE NO2
field calibration

SO2

SO2 MLE settings applied fw1out=(); fw2out=(); dqf2in=(); ddlim=(-1.000,-1.000); shilim(nm)=(-0.050,0.050); uscmax=1.163e-02; wrmsmax=9.000e-04; atvarmax(%)=8.000e+01; amflim=(1.000,3.000); amflimvar=(0.000,1.000); perc=0.00; ngroup=13; timeaver(min)=0.00; redu=0; scref=(7.627e-05); uscref=(1.272e-05)

FieldCalibration MLE SO2

Pandora104s1_FieldCalibration_MLE_SO2.jpeg - 104 KB
FieldCalibration MLE SO2

FieldCalibration VCMLE SO2

Pandora104s1_FieldCalibration_VCMLE_SO2.jpeg - 98 KB
FieldCalibration VCMLE SO2
field calibration

HCHO

HCHO redu=1 settings applied fw1out=(); fw2out=(); dqf2in=(); ddlim=(-1.000,-1.000); shilim(nm)=(-0.050,0.050); uscmax=1.179e-03; wrmsmax=9.000e-04; atvarmax(%)=5.000e+01; amflim=(1.000,5.000); amflimvar=(0.000,1.000); perc=0.00; ngroup=10; timeaver(min)=0.00; redu=1; scref=(3.448e-04); uscref=(7.999e-05)

FieldCalibration MLE HCHO

Pandora104s1_FieldCalibration_MLE_HCHO.jpeg - 105 KB
FieldCalibration MLE HCHO

FieldCalibration VCMLE HCHO

Pandora104s1_FieldCalibration_VCMLE_HCHO.jpeg - 108 KB
FieldCalibration VCMLE HCHO

Automatically harvested metadata

Automatically Harvested Metadata

Manual review and decision

Manual Review and Decision

Calibration notes, data-quality findings, and whole-period evaluation.

Data Evaluation

Could there be an offset between VPs?

No

Are there breakpoint-analysis plots or suspicious discontinuities?

No

Do daily figures before and after the new VP look consistent?

No

Is there an AMF dependency for one or more VPs?

No

Is there seasonality, for example negative values in winter?

No

Does the daily shape follow the trop data roughly?

No

If there is a discrepancy, what are the DS and Sky viewing directions?

No

Is there a co-located or nearby instrument for comparison?

No

O3Ext

Decision for O3Ext
With Care
Unit will need at least 3 more validity periods. Strong wl_shifts over the last years.

O3Synt

Decision for O3Synt
With Care

NO2

Decision for NO2
With Care
Unit will need at least 3 more validity periods. Strong wl_shifts over the last years.

SO2

Decision for SO2
With Care
Unit will need at least 3 more validity periods. Strong wl_shifts over the last years.

HCHO

Decision for HCHO
With Care
Unit will need at least 3 more validity periods. Strong wl_shifts over the last years.

H2O

Decision for H2O
Yes