Decision Summary
Upload-ready calibration protocol overview
Instrument
P104s1
Session
8
Location
Downsview
Calibration person
Ilias Bougoudis
CF file
Pandora104s1_CF_v8d20221101.txt
Software version
p1-8
Calibration vector
-
Reference day
20230529
Overall status
development
O3Ext: With CareO3Synt: With CareNO2: With CareSO2: With CareHCHO: With CareH2O: Yes
Action Timeline
60 action intervalsDuration by Step
Action Summary
| Action | Events | Runs | Aborted | Time |
|---|---|---|---|---|
| AssignCalFile | 3 | 1 | 0 | 43s |
| CreateReference | 53 | 27 | 9 | 7m 21s |
| FieldCalibration | 211 | 32 | 0 | 1h 51m |
| No action | 8 | 0 | 0 | 0s |
Session Figures
17 figures from Google Drive/figures
calprotocol section
Wavelength Calibration
reference period
Reference Day Decision Support
Upload the FOV and L0 raw-count evidence used to justify the selected reference day before reviewing the CreateReference figures.
calprotocol section
Reference Period and CreateReference
Create reference L1CHARFILT
Pandora104s1_CreateReference_L1CHARFILT_OPEN_U340.jpeg - 145 KBCreate reference SELSPECS
Pandora104s1_CreateReference_SELSPECS_OPEN_U340.jpeg - 117 KBCreate reference SLERR (STRAY)
Pandora104s1_CreateReference_SLERR(STRAY)_OPEN_U340.jpeg - 126 KBCreate reference AODEST
Pandora104s1_CreateReference_AODEST.jpeg - 54 KBCreate reference SENSCORRFIN (SENS)
Pandora104s1_CreateReference_SENSCORRFIN(SENS)_OPEN_U340.jpeg - 120 KBCreate reference DISPFIT (DISP)
Pandora104s1_CreateReference_DISPFIT(DISP).jpeg - 108 KBCreate reference WLCHANGE (DISP)
Pandora104s1_CreateReference_WLCHANGE(DISP)_OPEN_U340.jpeg - 116 KBCreate reference SHIFT (CLOS)
Pandora104s1_CreateReference_SHIFT(CLOS)_OPEN_U340.jpeg - 115 KB
field calibration
O3
field calibration
NO2
field calibration
SO2
field calibration
HCHO
Automatically harvested metadata
Automatically Harvested Metadata
Manual review and decision
Manual Review and Decision
Calibration notes, data-quality findings, and whole-period evaluation.Data Evaluation
Could there be an offset between VPs?
No
Are there breakpoint-analysis plots or suspicious discontinuities?
No
Do daily figures before and after the new VP look consistent?
No
Is there an AMF dependency for one or more VPs?
No
Is there seasonality, for example negative values in winter?
No
Does the daily shape follow the trop data roughly?
No
If there is a discrepancy, what are the DS and Sky viewing directions?
No
Is there a co-located or nearby instrument for comparison?
No
O3Ext
Decision for O3Ext
With Care
Unit will need at least 3 more validity periods. Strong wl_shifts over the last years.
O3Synt
Decision for O3Synt
With Care
NO2
Decision for NO2
With Care
Unit will need at least 3 more validity periods. Strong wl_shifts over the last years.
SO2
Decision for SO2
With Care
Unit will need at least 3 more validity periods. Strong wl_shifts over the last years.
HCHO
Decision for HCHO
With Care
Unit will need at least 3 more validity periods. Strong wl_shifts over the last years.
H2O
Decision for H2O
Yes
Unit will need at least 3 more validity periods. Strong wl_shifts over the last years.