Decision Summary
Upload-ready calibration protocol overview
Instrument
P104s1
Session
4
Location
Downsview
Calibration person
Ilias Bougoudis
CF file
Pandora104s1_CF_v4d20130709.txt
Software version
p1-8
Calibration vector
-
Reference day
-
Overall status
development
O3Ext: UnusableO3Synt: With CareNO2: YesSO2: With CareHCHO: With CareH2O: With Care
Action Timeline
30 action intervalsDuration by Step
Action Summary
| Action | Events | Runs | Aborted | Time |
|---|---|---|---|---|
| AssignCalFile | 3 | 1 | 0 | 27s |
| FieldCalibration | 205 | 29 | 3 | 2h 13m |
| No action | 12 | 0 | 0 | 0s |
Session Figures
No session figures found for the loaded instrument/session.No figures loaded
Automatically harvested metadata
Automatically Harvested Metadata
Manual review and decision
Manual Review and Decision
Calibration notes, data-quality findings, and whole-period evaluation.Data Evaluation
Could there be an offset between VPs?
No
Are there breakpoint-analysis plots or suspicious discontinuities?
No
Do daily figures before and after the new VP look consistent?
No
Is there an AMF dependency for one or more VPs?
No
Is there seasonality, for example negative values in winter?
No
Does the daily shape follow the trop data roughly?
No
If there is a discrepancy, what are the DS and Sky viewing directions?
No
Is there a co-located or nearby instrument for comparison?
No
O3Ext
Decision for O3Ext
Unusable
O3Ext unusable again.
O3Synt
Decision for O3Synt
With Care
NO2
Decision for NO2
Yes
good over all periods
SO2
Decision for SO2
With Care
good until end of summer 2021
HCHO
Decision for HCHO
With Care
good until end of summer 2021
H2O
Decision for H2O
With Care
flagged out over the early days