Decision Summary

Upload-ready calibration protocol overview
Instrument P104s1
Session 22
Location -
Calibration person Ragi Rajagopalan
CF file Pandora104s1_CF_v22d20211217.txt
Software version 1.8.66
Calibration vector -
Reference day -
Overall status development
O3Ext: Not answeredO3Synt: Not answeredNO2: Not answeredSO2: YesHCHO: Not answeredH2O: Not answered
Instrument P104s1
Session 22
Events 20
Figures 17
Action time 9m 46s

Action Timeline

5 action intervals
Nov 25, 202403:59 PM04:25 PMFieldCalibrationAssignCalFileFieldCalibration 20s abortedFieldCalibration 3m 10s abortedFieldCalibration 5m 27s finishedFieldCalibration 0s abortedAssignCalFile 49s finished

Duration by Step

FieldCalibration8m 57sAssignCalFile49s

Action Summary

Action Events Runs Aborted Time
AssignCalFile 3 1 0 49s
FieldCalibration 8 4 3 8m 57s
No action 9 0 0 0s

Session Figures

17 figures from Google Drive/figures
calprotocol section

Wavelength Calibration

Slit function parameters

Pandora104s1_WavelengthCalibration_SlitFuncParam.jpeg - 153 KB
Slit function parameters
reference period

Reference Day Decision Support

Upload the FOV and L0 raw-count evidence used to justify the selected reference day before reviewing the CreateReference figures.

calprotocol section

Reference Period and CreateReference

Create reference L1CHARFILT

Pandora104s1_CreateReference_L1CHARFILT_OPEN_U340.jpeg - 149 KB
Create reference L1CHARFILT

Create reference SELSPECS

Pandora104s1_CreateReference_SELSPECS_OPEN_U340.jpeg - 128 KB
Create reference SELSPECS

Create reference SLERR (STRAY)

Pandora104s1_CreateReference_SLERR(STRAY)_OPEN_U340.jpeg - 120 KB
Create reference SLERR (STRAY)

Create reference AODEST

Pandora104s1_CreateReference_AODEST.jpeg - 68 KB
Create reference AODEST

Create reference SENSCORRFIN (SENS)

Pandora104s1_CreateReference_SENSCORRFIN(SENS)_OPEN_U340.jpeg - 118 KB
Create reference SENSCORRFIN (SENS)

Create reference DISPFIT (DISP)

Pandora104s1_CreateReference_DISPFIT(DISP).jpeg - 109 KB
Create reference DISPFIT (DISP)

Create reference WLCHANGE (DISP)

Pandora104s1_CreateReference_WLCHANGE(DISP)_OPEN_U340.jpeg - 194 KB
Create reference WLCHANGE (DISP)

Create reference SHIFT (CLOS)

Pandora104s1_CreateReference_SHIFT(CLOS)_OPEN_U340.jpeg - 155 KB
Create reference SHIFT (CLOS)
field calibration

O3

FieldCalibration VC O3

Pandora104s1_FieldCalibration_VC_O3.jpeg - 126 KB
FieldCalibration VC O3

FieldCalibration SCREFINT O3

Pandora104s1_FieldCalibration_SCREFINT_O3.jpeg - 81 KB
FieldCalibration SCREFINT O3
field calibration

NO2

FieldCalibration MLE NO2

Pandora104s1_FieldCalibration_MLE_NO2.jpeg - 171 KB
FieldCalibration MLE NO2

FieldCalibration VCMLE NO2

Pandora104s1_FieldCalibration_VCMLE_NO2.jpeg - 106 KB
FieldCalibration VCMLE NO2
field calibration

SO2

SO2 MLE settings applied fw1out=(); fw2out=(); dqf2in=(103); ddlim=(-1.000,-1.000); shilim(nm)=(-0.050,0.050); uscmax=9.900e-04; wrmsmax=9.000e-04; atvarmax(%)=1.000e+02; amflim=(1.000,2.000); amflimvar=(0.000,1.000); perc=10.00; ngroup=10; timeaver(min)=0.00; redu=0; scref=(1.416e-05); uscref=(1.129e-05)

FieldCalibration MLE SO2

Pandora104s1_FieldCalibration_MLE_SO2.jpeg - 64 KB
FieldCalibration MLE SO2

FieldCalibration VCMLE SO2

Pandora104s1_FieldCalibration_VCMLE_SO2.jpeg - 60 KB
FieldCalibration VCMLE SO2
field calibration

HCHO

FieldCalibration MLE HCHO

Pandora104s1_FieldCalibration_MLE_HCHO.jpeg - 115 KB
FieldCalibration MLE HCHO

FieldCalibration VCMLE HCHO

Pandora104s1_FieldCalibration_VCMLE_HCHO.jpeg - 109 KB
FieldCalibration VCMLE HCHO

Automatically harvested metadata

Automatically Harvested Metadata

Manual review and decision

Manual Review and Decision

Calibration notes, data-quality findings, and whole-period evaluation.

Data Evaluation

Could there be an offset between VPs?

No

Are there breakpoint-analysis plots or suspicious discontinuities?

No

Do daily figures before and after the new VP look consistent?

No

Is there an AMF dependency for one or more VPs?

No

Is there seasonality, for example negative values in winter?

No

Does the daily shape follow the trop data roughly?

No

If there is a discrepancy, what are the DS and Sky viewing directions?

No

Is there a co-located or nearby instrument for comparison?

No

O3Ext

Decision for O3Ext
No change on O3

O3Synt

Decision for O3Synt

NO2

Decision for NO2
No change on NO2

SO2

Decision for SO2
Yes
Redid SO2 MLE

HCHO

Decision for HCHO
No calibration on HCHO delrin

H2O

Decision for H2O