Decision Summary

Upload-ready calibration protocol overview
Instrument P104s1
Session 12
Location Aldino
Calibration person Ragi Rajagopalan
CF file Pandora104s1_CF_v12d20180627.txt
Software version 1.8.57
Calibration vector -
Reference day 20180811
Overall status development
O3Ext: With CareO3Synt: With CareNO2: YesSO2: YesHCHO: Not answeredH2O: Not answered
Instrument P104s1
Session 12
Events 464
Figures 17
Action time 2h 17m

Action Timeline

94 action intervals
Aug 02, 2023Aug 03, 202308:04 AM12:38 PMCheckFiltersCheckNoiseWaveAnalyzeDarkAnalyzeBrightWavelengthCalibrationTempDependenceNDFiltersSensitivityCalibrationConvolutionCreateReferenceFieldCalibrationAssignCalFileCheckFilters 6s finishedCheckNoiseWave 9s finishedAnalyzeDark 1m 39s finishedAnalyzeBright 1m 59s finishedWavelengthCalibration 5m 39s finishedWavelengthCalibration 0s finishedWavelengthCalibration 0s finishedWavelengthCalibration 0s finishedWavelengthCalibration 0s finishedTempDependence 34s finishedNDFilters 3m 26s finishedNDFilters 2m 21s finishedSensitivityCalibration 2m 44s finishedSensitivityCalibration 0s finishedSensitivityCalibration 0s finishedConvolution 52s finishedConvolution 0s finishedConvolution 0s finishedConvolution 0s finishedConvolution 0s finishedCreateReference 25m 44s abortedCreateReference 15s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 40s finishedCreateReference 0s abortedCreateReference 2m 13s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 3m 53s abortedCreateReference 31s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 13s finishedCreateReference 0s abortedCreateReference 56s finishedCreateReference 0s abortedWavelengthCalibration 6s abortedWavelengthCalibration 12s finishedCreateReference 16s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 5s abortedCreateReference 49s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 1m 35s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 27s abortedCreateReference 31s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 23s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 9s finishedCreateReference 0s finishedFieldCalibration 2m 3s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 52m 24s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 13m 7s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 10m 30s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedAssignCalFile 31s finished

Duration by Step

FieldCalibration1h 18mCreateReference38m 41sWavelengthCalibration5m 57sNDFilters5m 48sSensitivityCalibration2m 44sAnalyzeBright1m 59sAnalyzeDark1m 39sConvolution52sTempDependence34sAssignCalFile31sCheckNoiseWave9sCheckFilters6s

Action Summary

Action Events Runs Aborted Time
AnalyzeBright 12 1 0 1m 59s
AnalyzeDark 9 1 0 1m 39s
AssignCalFile 3 1 0 31s
CheckFilters 3 1 0 6s
CheckNoiseWave 5 1 0 9s
Convolution 16 5 0 52s
CreateReference 87 41 15 38m 41s
FieldCalibration 155 30 0 1h 18m
NDFilters 66 2 0 5m 48s
No action 15 0 0 0s
SensitivityCalibration 64 3 0 2m 44s
TempDependence 6 1 0 34s
WavelengthCalibration 23 7 1 5m 57s

Session Figures

17 figures from Google Drive/figures
calprotocol section

Wavelength Calibration

Slit function parameters

Pandora104s1_WavelengthCalibration_SlitFuncParam.jpeg - 153 KB
Slit function parameters
reference period

Reference Day Decision Support

Upload the FOV and L0 raw-count evidence used to justify the selected reference day before reviewing the CreateReference figures.

calprotocol section

Reference Period and CreateReference

Create reference L1CHARFILT

Pandora104s1_CreateReference_L1CHARFILT_OPEN_U340.jpeg - 152 KB
Create reference L1CHARFILT

Create reference SELSPECS

Pandora104s1_CreateReference_SELSPECS_OPEN_U340.jpeg - 137 KB
Create reference SELSPECS

Create reference SLERR (STRAY)

Pandora104s1_CreateReference_SLERR(STRAY)_OPEN_U340.jpeg - 125 KB
Create reference SLERR (STRAY)

Create reference AODEST

Pandora104s1_CreateReference_AODEST.jpeg - 67 KB
Create reference AODEST

Create reference SENSCORRFIN (SENS)

Pandora104s1_CreateReference_SENSCORRFIN(SENS)_OPEN_U340.jpeg - 115 KB
Create reference SENSCORRFIN (SENS)

Create reference DISPFIT (DISP)

Pandora104s1_CreateReference_DISPFIT(DISP).jpeg - 101 KB
Create reference DISPFIT (DISP)

Create reference WLCHANGE (DISP)

Pandora104s1_CreateReference_WLCHANGE(DISP)_OPEN_U340.jpeg - 196 KB
Create reference WLCHANGE (DISP)

Create reference SHIFT (CLOS)

Pandora104s1_CreateReference_SHIFT(CLOS)_OPEN_U340.jpeg - 133 KB
Create reference SHIFT (CLOS)
field calibration

O3

FieldCalibration VC O3

Pandora104s1_FieldCalibration_VC_O3.jpeg - 113 KB
FieldCalibration VC O3

FieldCalibration SCREFINT O3

Pandora104s1_FieldCalibration_SCREFINT_O3.jpeg - 77 KB
FieldCalibration SCREFINT O3
field calibration

NO2

NO2 redu=1 settings applied fw1out=(); fw2out=(); dqf2in=(); ddlim=(-1.000,-1.000); shilim(nm)=(-0.050,0.050); uscmax=4.893e-03; wrmsmax=1.774e+00; atvarmax(%)=2.000e+01; amflim=(1.000,5.000); amflimvar=(0.000,1.000); perc=0.00; ngroup=30; timeaver(min)=0.00; redu=1; scref=(2.248e-04,1.907e-04,1.857e-04); uscref=(1.853e-05,7.914e-06,1.792e-05)

FieldCalibration MLE NO2

Pandora104s1_FieldCalibration_MLE_NO2.jpeg - 144 KB
FieldCalibration MLE NO2

FieldCalibration VCMLE NO2

Pandora104s1_FieldCalibration_VCMLE_NO2.jpeg - 90 KB
FieldCalibration VCMLE NO2
field calibration

SO2

SO2 MLE settings applied fw1out=(); fw2out=(); dqf2in=(); ddlim=(-1.000,-1.000); shilim(nm)=(-0.050,0.050); uscmax=2.553e-02; wrmsmax=1.552e+00; atvarmax(%)=5.000e+01; amflim=(1.000,2.500); amflimvar=(0.000,1.000); perc=0.00; ngroup=10; timeaver(min)=0.00; redu=0; scref=(1.243e-04); uscref=(5.716e-05)

FieldCalibration MLE SO2

Pandora104s1_FieldCalibration_MLE_SO2.jpeg - 101 KB
FieldCalibration MLE SO2

FieldCalibration VCMLE SO2

Pandora104s1_FieldCalibration_VCMLE_SO2.jpeg - 91 KB
FieldCalibration VCMLE SO2
field calibration

HCHO

HCHO MLE settings applied fw1out=(); fw2out=(); dqf2in=(); ddlim=(-1.000,-1.000); shilim(nm)=(-0.050,0.050); uscmax=4.274e-02; wrmsmax=1.788e+00; atvarmax(%)=5.000e+01; amflim=(1.000,5.000); amflimvar=(0.000,1.000); perc=0.00; ngroup=10; timeaver(min)=0.00; redu=0; scref=(2.087e-04); uscref=(4.787e-05)

FieldCalibration MLE HCHO

Pandora104s1_FieldCalibration_MLE_HCHO.jpeg - 107 KB
FieldCalibration MLE HCHO

FieldCalibration VCMLE HCHO

Pandora104s1_FieldCalibration_VCMLE_HCHO.jpeg - 95 KB
FieldCalibration VCMLE HCHO

Automatically harvested metadata

Automatically Harvested Metadata

Manual review and decision

Manual Review and Decision

Calibration notes, data-quality findings, and whole-period evaluation.

Data Evaluation

Could there be an offset between VPs?

No

Are there breakpoint-analysis plots or suspicious discontinuities?

No

Do daily figures before and after the new VP look consistent?

No

Is there an AMF dependency for one or more VPs?

No

Is there seasonality, for example negative values in winter?

No

Does the daily shape follow the trop data roughly?

No

If there is a discrepancy, what are the DS and Sky viewing directions?

No

Is there a co-located or nearby instrument for comparison?

No

O3Ext

Decision for O3Ext
With Care
Looks good but W-shape in wrms somedays

O3Synt

Decision for O3Synt
With Care

NO2

Decision for NO2
Yes
good over all periods

SO2

Decision for SO2
Yes
Looks good

HCHO

Decision for HCHO
no calibration still has delrin

H2O

Decision for H2O