Decision Summary
Upload-ready calibration protocol overview
Instrument
P103s1
Session
9
Location
Downsview
Calibration person
Hanna Rohringer
CF file
Pandora103s1_CF_v9d20200420.txt
Software version
p1-8
Calibration vector
-
Reference day
20200607
Overall status
development
O3Ext: With CareO3Synt: With CareNO2: YesSO2: YesHCHO: UnusableH2O: With Care
Action Timeline
127 action intervalsDuration by Step
Action Summary
| Action | Events | Runs | Aborted | Time |
|---|---|---|---|---|
| AssignCalFile | 3 | 1 | 0 | 3m 11s |
| CreateReference | 143 | 62 | 18 | 1h 32m |
| FieldCalibration | 244 | 59 | 15 | 3h 50m |
| No action | 24 | 0 | 0 | 0s |
| SensitivityCalibration | 38 | 3 | 1 | 21s |
| WavelengthCalibration | 6 | 2 | 0 | 26s |
Session Figures
No session figures found for the loaded instrument/session.No figures loaded
Automatically harvested metadata
Automatically Harvested Metadata
Manual review and decision
Manual Review and Decision
Calibration notes, data-quality findings, and whole-period evaluation.Data Evaluation
Could there be an offset between VPs?
No
Are there breakpoint-analysis plots or suspicious discontinuities?
No
Do daily figures before and after the new VP look consistent?
No
Is there an AMF dependency for one or more VPs?
No
Is there seasonality, for example negative values in winter?
No
Does the daily shape follow the trop data roughly?
No
If there is a discrepancy, what are the DS and Sky viewing directions?
No
Is there a co-located or nearby instrument for comparison?
No
O3Ext
Decision for O3Ext
With Care
AXC-Temperature significantly higher than climatology (+14°C), but shape of Temperature and vc look ok, different fiber in the lab as in the field - bad out
O3Synt
Decision for O3Synt
With Care
NO2
Decision for NO2
Yes
look spiky but no AMF dependecie
SO2
Decision for SO2
Yes
look good
HCHO
Decision for HCHO
Unusable
Delrin
H2O
Decision for H2O
With Care
mostly wrms flagged, different fiber in the lab as in the field - bad out