Decision Summary
Upload-ready calibration protocol overview
Instrument
P103s1
Session
6
Location
Downsview
Calibration person
Trainee
CF file
Pandora103s1_CF_v6d20230525.txt
Software version
1.8.60
Calibration vector
-
Reference day
20230620
Overall status
development
O3Ext: With CareO3Synt: With CareNO2: YesSO2: With CareHCHO: With CareH2O: With Care
Action Timeline
207 action intervalsDuration by Step
Action Summary
| Action | Events | Runs | Aborted | Time |
|---|---|---|---|---|
| AssignCalFile | 3 | 1 | 0 | 1m 22s |
| Convolution | 32 | 10 | 0 | 2m 1s |
| CreateReference | 111 | 54 | 19 | 2h 10m |
| FieldCalibration | 521 | 137 | 48 | 5h 56m |
| NDFilters | 2 | 1 | 1 | 1m 10s |
| No action | 62 | 0 | 0 | 0s |
| WavelengthCalibration | 12 | 4 | 0 | 58s |
Session Figures
17 figures from Google Drive/figures
calprotocol section
Wavelength Calibration
reference period
Reference Day Decision Support
Upload the FOV and L0 raw-count evidence used to justify the selected reference day before reviewing the CreateReference figures.
calprotocol section
Reference Period and CreateReference
Create reference L1CHARFILT
Pandora103s1_CreateReference_L1CHARFILT_OPEN_U340.jpeg - 241 KBCreate reference SELSPECS
Pandora103s1_CreateReference_SELSPECS_OPEN_U340.jpeg - 228 KBCreate reference SLERR (STRAY)
Pandora103s1_CreateReference_SLERR(STRAY)_OPEN_U340.jpeg - 178 KBCreate reference AODEST
Pandora103s1_CreateReference_AODEST.jpeg - 95 KBCreate reference SENSCORRFIN (SENS)
Pandora103s1_CreateReference_SENSCORRFIN(SENS)_OPEN_U340.jpeg - 154 KBCreate reference DISPFIT (DISP)
Pandora103s1_CreateReference_DISPFIT(DISP).jpeg - 73 KBCreate reference WLCHANGE (DISP)
Pandora103s1_CreateReference_WLCHANGE(DISP)_OPEN_U340.jpeg - 125 KBCreate reference SHIFT (CLOS)
Pandora103s1_CreateReference_SHIFT(CLOS)_OPEN_U340.jpeg - 218 KB
field calibration
O3
field calibration
NO2
field calibration
SO2
field calibration
HCHO
Automatically harvested metadata
Automatically Harvested Metadata
Manual review and decision
Manual Review and Decision
Calibration notes, data-quality findings, and whole-period evaluation.Data Evaluation
Could there be an offset between VPs?
No
Are there breakpoint-analysis plots or suspicious discontinuities?
No
Do daily figures before and after the new VP look consistent?
No
Is there an AMF dependency for one or more VPs?
No
Is there seasonality, for example negative values in winter?
No
Does the daily shape follow the trop data roughly?
No
If there is a discrepancy, what are the DS and Sky viewing directions?
No
Is there a co-located or nearby instrument for comparison?
No
O3Ext
Decision for O3Ext
With Care
Used Lorenzian as for gaussian AXC-T was significantly higher then clim
O3Synt
Decision for O3Synt
With Care
NO2
Decision for NO2
Yes
SO2
Decision for SO2
With Care
same as for HCHO
HCHO
Decision for HCHO
With Care
mostly flagged, in teamwork with CW we tried to make it better (use Gaussian + ClimT for o3) but did not achieve any success. Anticpate the reason is the bad stray light which cant be taken into account in 1.8
H2O
Decision for H2O
With Care