Decision Summary
Upload-ready calibration protocol overview
Instrument
P100s1
Session
5
Location
NewHavenCT
Calibration person
Alexander Kotsakis
CF file
Pandora100s1_CF_v5d20240611.txt
Software version
1.8.88
Calibration vector
CalV1
Reference day
20250809
Overall status
development
O3Ext: With CareO3Synt: With CareNO2: YesSO2: YesHCHO: YesH2O: Yes
Action Timeline
225 action intervalsDuration by Step
Action Summary
| Action | Events | Runs | Aborted | Time |
|---|---|---|---|---|
| AssignCalFile | 13 | 4 | 1 | 10m 49s |
| CheckDispersion | 102 | 33 | 1 | 22m 46s |
| CheckFilters | 4 | 1 | 0 | 12s |
| CheckNoiseWave | 6 | 1 | 0 | 2m 16s |
| Convolution | 16 | 5 | 0 | 38s |
| CreateReference | 297 | 73 | 34 | 9h 4m |
| FieldCalibration | 439 | 77 | 30 | 19h 25m |
| NDFilters | 60 | 10 | 9 | 6m 1s |
| No action | 83 | 0 | 0 | 0s |
| SensitivityCalibration | 29 | 2 | 0 | 5m 28s |
| StrayLight | 5 | 2 | 2 | 8s |
Session Figures
17 figures from Google Drive/figures
calprotocol section
Wavelength Calibration
reference period
Reference Day Decision Support
Upload the FOV and L0 raw-count evidence used to justify the selected reference day before reviewing the CreateReference figures.
calprotocol section
Reference Period and CreateReference
Create reference L1CHARFILT
Pandora100s1_CreateReference_L1CHARFILT_OPEN_U340.jpeg - 168 KBCreate reference SELSPECS
Pandora100s1_CreateReference_SELSPECS_OPEN_U340.jpeg - 119 KBCreate reference SLERR (STRAY)
Pandora100s1_CreateReference_SLERR(STRAY)_OPEN_U340.jpeg - 200 KBCreate reference AODEST
Pandora100s1_CreateReference_AODEST.jpeg - 62 KBCreate reference SENSCORRFIN (SENS)
Pandora100s1_CreateReference_SENSCORRFIN(SENS)_OPEN_U340.jpeg - 118 KBCreate reference DISPFIT (DISP)
Pandora100s1_CreateReference_DISPFIT(DISP).jpeg - 111 KBCreate reference WLCHANGE (DISP)
Pandora100s1_CreateReference_WLCHANGE(DISP)_OPEN_U340.jpeg - 115 KBCreate reference SHIFT (CLOS)
Pandora100s1_CreateReference_SHIFT(CLOS)_OPEN_U340.jpeg - 131 KB
field calibration
O3
field calibration
NO2
field calibration
SO2
field calibration
HCHO
Automatically harvested metadata
Automatically Harvested Metadata
Manual review and decision
Manual Review and Decision
Calibration notes, data-quality findings, and whole-period evaluation.Data Evaluation
Could there be an offset between VPs?
No
Are there breakpoint-analysis plots or suspicious discontinuities?
No
Do daily figures before and after the new VP look consistent?
No
Is there an AMF dependency for one or more VPs?
No
Is there seasonality, for example negative values in winter?
No
Does the daily shape follow the trop data roughly?
No
If there is a discrepancy, what are the DS and Sky viewing directions?
No
Is there a co-located or nearby instrument for comparison?
No
O3Ext
Decision for O3Ext
With Care
out2 mostly flagged out. O3Temp has inverse-u shape
O3Synt
Decision for O3Synt
With Care
NO2
Decision for NO2
Yes
looks good, E-MLE used
SO2
Decision for SO2
Yes
looks good
HCHO
Decision for HCHO
Yes
looks good
H2O
Decision for H2O
Yes
looks good