Decision Summary

Upload-ready calibration protocol overview
Instrument P52s1
Session 5
Location LabGSFC
Calibration person Fernando Santos
CF file Pandora52s1_CF_v4d20170506.txt
Software version 1.7
Calibration vector -
Reference day -
Overall status development
O3Ext: Not answeredO3Synt: Not answeredNO2: Not answeredSO2: Not answeredHCHO: Not answeredH2O: Not answered
Instrument P52s1
Session 5
Events 204
Figures 1
Action time 2h 46m

Action Timeline

31 action intervals
Oct 03, 2019Oct 11, 2019Oct 11, 201902:54 PMSensitivityCalibrationConvolutionCheckDispersionCreateReferenceAssignCalFileSensitivityCalibration 18m 34s finishedSensitivityCalibration 0s finishedConvolution 5m 53s finishedConvolution 0s finishedConvolution 0s finishedConvolution 0s finishedConvolution 1m 55s finishedConvolution 0s finishedConvolution 0s finishedConvolution 0s finishedConvolution 0s finishedCheckDispersion 54m 11s finishedCheckDispersion 48s abortedCheckDispersion 44s abortedCheckDispersion 27m 33s finishedCheckDispersion 3s finishedCheckDispersion 1m 55s finishedCheckDispersion 3s finishedCheckDispersion 0s finishedCreateReference 12m 6s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 5m 6s finishedCreateReference 0s finishedCreateReference 0s finishedAssignCalFile 3m 9s finishedAssignCalFile 15m 55s abortedAssignCalFile 4m 50s abortedAssignCalFile 21s abortedAssignCalFile 13m 1s finishedAssignCalFile 0s open

Duration by Step

CheckDispersion1h 25mAssignCalFile37m 16sSensitivityCalibration18m 34sCreateReference17m 11sConvolution7m 47s

Action Summary

Action Events Runs Aborted Time
AssignCalFile 17 5 3 37m 16s
CheckDispersion 25 8 2 1h 25m
Convolution 32 9 0 7m 47s
CreateReference 86 6 0 17m 11s
No action 15 0 0 0s
SensitivityCalibration 29 2 0 18m 34s

Session Figures

1 figure from Google Drive/figures
calprotocol section

Wavelength Calibration

Slit function parameters

Pandora52s1_WavelengthCalibration_SlitFuncParam(A2).jpeg - 78 KB
Slit function parameters

Automatically harvested metadata

Automatically Harvested Metadata

Manual review and decision

Manual Review and Decision

Calibration notes, data-quality findings, and whole-period evaluation.

Data Evaluation

Could there be an offset between VPs?

No

Are there breakpoint-analysis plots or suspicious discontinuities?

No

Do daily figures before and after the new VP look consistent?

No

Is there an AMF dependency for one or more VPs?

No

Is there seasonality, for example negative values in winter?

No

Does the daily shape follow the trop data roughly?

No

If there is a discrepancy, what are the DS and Sky viewing directions?

No

Is there a co-located or nearby instrument for comparison?

No

O3Ext

Decision for O3Ext

O3Synt

Decision for O3Synt

NO2

Decision for NO2

SO2

Decision for SO2

HCHO

Decision for HCHO

H2O

Decision for H2O