Decision Summary

Upload-ready calibration protocol overview
Instrument P40s1
Session 12
Location LabGSFC
Calibration person Leonie Haunold
CF file C:/Blick-1.8-58/data/calresults/Pandora40s1/session12/Pandora40s1_CF_session12.txt
Software version 1.7
Calibration vector -
Reference day -
Overall status development
O3Ext: Not answeredO3Synt: Not answeredNO2: Not answeredSO2: Not answeredHCHO: Not answeredH2O: Not answered
Instrument P40s1
Session 12
Events 130
Figures 11
Action time 14m 4s

Action Timeline

17 action intervals
Mar 04, 202501:56 PM03:42 PMWavelengthCalibrationTempDependenceNDFiltersSensitivityCalibrationAnalyzeBrightConvolutionAssignCalFileWavelengthCalibration 7m 44s finishedTempDependence 1m 39s abortedTempDependence 30s finishedTempDependence 19s finishedNDFilters 22s finishedSensitivityCalibration 11s abortedNDFilters 9s abortedAnalyzeBright 23s abortedSensitivityCalibration 33s finishedSensitivityCalibration 0s finishedSensitivityCalibration 0s finishedConvolution 2m 5s finishedConvolution 0s finishedConvolution 0s finishedConvolution 0s finishedConvolution 0s finishedAssignCalFile 11s aborted

Duration by Step

WavelengthCalibration7m 44sTempDependence2m 28sConvolution2m 5sSensitivityCalibration44sNDFilters31sAnalyzeBright23sAssignCalFile11s

Action Summary

Action Events Runs Aborted Time
AnalyzeBright 3 1 1 23s
AssignCalFile 2 1 1 11s
Convolution 16 5 0 2m 5s
NDFilters 16 2 1 31s
No action 9 0 0 0s
SensitivityCalibration 66 4 1 44s
TempDependence 15 3 1 2m 28s
WavelengthCalibration 3 1 0 7m 44s

Session Figures

11 figures from Google Drive/figures
calprotocol section

Wavelength Calibration

Slit function parameters

Pandora40s1_WavelengthCalibration_SlitFuncParam.jpeg - 118 KB
Slit function parameters
reference period

Reference Day Decision Support

Upload the FOV and L0 raw-count evidence used to justify the selected reference day before reviewing the CreateReference figures.

calprotocol section

Reference Period and CreateReference

Create reference L1CHARFILT

Pandora40s1_CreateReference_L1CHARFILT_OPEN_U340.jpeg - 84 KB
Create reference L1CHARFILT

Create reference SELSPECS

Pandora40s1_CreateReference_SELSPECS_OPEN_U340.jpeg - 73 KB
Create reference SELSPECS

Create reference SLERR (STRAY)

Pandora40s1_CreateReference_SLERR(STRAY)_OPEN_U340.jpeg - 72 KB
Create reference SLERR (STRAY)

Create reference AODEST

Pandora40s1_CreateReference_AODEST.jpeg - 32 KB
Create reference AODEST

Create reference SENSCORRFIN (SENS)

Pandora40s1_CreateReference_SENSCORRFIN(SENS)_OPEN_U340.jpeg - 73 KB
Create reference SENSCORRFIN (SENS)

Create reference DISPFIT (DISP)

Pandora40s1_CreateReference_DISPFIT(DISP).jpeg - 58 KB
Create reference DISPFIT (DISP)

Create reference WLCHANGE (DISP)

Pandora40s1_CreateReference_WLCHANGE(DISP)_OPEN_U340.jpeg - 105 KB
Create reference WLCHANGE (DISP)

Create reference SHIFT (CLOS)

Pandora40s1_CreateReference_SHIFT(CLOS)_OPEN_U340.jpeg - 87 KB
Create reference SHIFT (CLOS)
field calibration

O3

FieldCalibration VC O3

Pandora40s1_FieldCalibration_VC_O3.jpeg - 59 KB
FieldCalibration VC O3

FieldCalibration SCREFINT O3

Pandora40s1_FieldCalibration_SCREFINT_O3.jpeg - 46 KB
FieldCalibration SCREFINT O3

Automatically harvested metadata

Automatically Harvested Metadata

Manual review and decision

Manual Review and Decision

Calibration notes, data-quality findings, and whole-period evaluation.

Data Evaluation

Could there be an offset between VPs?

No

Are there breakpoint-analysis plots or suspicious discontinuities?

No

Do daily figures before and after the new VP look consistent?

No

Is there an AMF dependency for one or more VPs?

No

Is there seasonality, for example negative values in winter?

No

Does the daily shape follow the trop data roughly?

No

If there is a discrepancy, what are the DS and Sky viewing directions?

No

Is there a co-located or nearby instrument for comparison?

No

O3Ext

Decision for O3Ext

O3Synt

Decision for O3Synt

NO2

Decision for NO2

SO2

Decision for SO2

HCHO

Decision for HCHO

H2O

Decision for H2O