Decision Summary

Upload-ready calibration protocol overview
Instrument P144s1
Session 5
Location Downsview
Calibration person Manuel Gebetsberger
CF file Pandora144s1_CF_v1d20191119.txt
Software version 1.7
Calibration vector -
Reference day -
Overall status development
O3Ext: Not answeredO3Synt: Not answeredNO2: Not answeredSO2: Not answeredHCHO: Not answeredH2O: Not answered
Instrument P144s1
Session 5
Events 155
Figures 1
Action time 39m 13s

Action Timeline

34 action intervals
May 29, 2020Jun 02, 202001:54 PM12:25 PMCreateReferenceCheckDispersionAssignCalFileCreateReference 2s abortedCheckDispersion 4s abortedCheckDispersion 0s staleCheckDispersion 0s staleCheckDispersion 3m 0s abortedCheckDispersion 0s staleCheckDispersion 46s abortedCheckDispersion 0s staleCheckDispersion 0s staleCheckDispersion 3m 17s abortedCheckDispersion 0s staleCheckDispersion 4s abortedCheckDispersion 34s finishedCheckDispersion 6s finishedCheckDispersion 0s finishedCheckDispersion 37s finishedCheckDispersion 1m 51s finishedCheckDispersion 5s finishedCheckDispersion 0s finishedCreateReference 1m 53s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 24m 17s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 1m 19s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedAssignCalFile 1m 20s finished

Duration by Step

CreateReference27m 30sCheckDispersion10m 23sAssignCalFile1m 20s

Action Summary

Action Events Runs Aborted Time
AssignCalFile 3 1 0 1m 20s
CheckDispersion 44 12 5 10m 23s
CreateReference 100 15 4 27m 30s
No action 8 0 0 0s

Session Figures

1 figure from Google Drive/figures
calprotocol section

Wavelength Calibration

Slit function parameters

Pandora144s1_WavelengthCalibration_SlitFuncParam(A2).jpeg - 46 KB
Slit function parameters

Automatically harvested metadata

Automatically Harvested Metadata

Manual review and decision

Manual Review and Decision

Calibration notes, data-quality findings, and whole-period evaluation.

Data Evaluation

Could there be an offset between VPs?

No

Are there breakpoint-analysis plots or suspicious discontinuities?

No

Do daily figures before and after the new VP look consistent?

No

Is there an AMF dependency for one or more VPs?

No

Is there seasonality, for example negative values in winter?

No

Does the daily shape follow the trop data roughly?

No

If there is a discrepancy, what are the DS and Sky viewing directions?

No

Is there a co-located or nearby instrument for comparison?

No

O3Ext

Decision for O3Ext

O3Synt

Decision for O3Synt

NO2

Decision for NO2

SO2

Decision for SO2

HCHO

Decision for HCHO

H2O

Decision for H2O